记录验证系统中床位容差表的设定方法
黄鹏 , 胡志辉 , 符贵山 , 张寅 , 覃仕瑞 , 戴建荣
国家癌症中心/国家肿瘤临床医学研究中心/中国医学科学院北京协和医学院肿瘤医院放疗科100021
A setting approach for tolerance table of couch position in record and verify system
Huang Peng, Hu Zhihui, Fu Guishan, Zhang Yin, Qin Shirui, Dai Jianrong
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences ( CAMS ) and Peking Union Medical College ( PUMC)
Abstract:Objective To evaluate the effect of the tolerance table of different couch positions in the record and verify system (R&V system) upon the setup accuracy.Methods Clinical data of 715 patients (15743 fractions of treatment) were extracted from the R&V system database and assigned into four categories including thorax and abdomen,head and neck,breast,and rectum based on the disease site and immobilization device. The first day couch position (FstD) and cumulative average couch position (CumA) were utilized as the references to analyze the couch setup of each faction of treatment,and to establish the tolerance tables of different sites. The sensitivity and specificity of two Methods were evaluated by the actual clinical treatment record of the patients.Results For the FstD as the reference,the couch tolerance in the breast was significantly higher than those in other parts. When the CumA was used as the reference,the couch tolerance tended to be stable after a certain fractions of treatment,and the tolerance of all sites was less than that of the FstD.The tolerance tables significantly differed between these two Methods (P=0.000).Both two Methods possessed high specificity,whereas the CumA method yielded higher sensitivity than the FstD approach.Conclusions Setting a reasonable tolerance table of couch position can effectively improve the setup accuracy.
Huang Peng,Hu Zhihui,Fu Guishan et al. A setting approach for tolerance table of couch position in record and verify system[J]. Chinese Journal of Radiation Oncology, 2019, 28(5): 373-377.
[1]Patton GA,Gaffney DK,Moeller JH.Facilitation of radiotherapeutic error by computerized record and verify systems[J].Int J Radiat Oncol Biol Phys,2003,56(1):50-57.DOI:10.1016/s0360-3016(02)04418-8. [2]Ford EC,Terezakis S,Souranis A,et al. Quality control quantification (QCQ):A tool to measure the value of quality control checks in radiation oncology[J].Int J Radiat Oncol Biol Phys,2012,84(3):263-269.DOI:10.1016/j.ijrobp.2012.04.036. [3]Yan G,Mittauer K,Huang Y,et al. Prevention of gross setup errors in radiotherapy with an efficient automatic patient safety system[J].J Appl Clin Med Phys,2013,14(6):322-337.DOI:10.1120/jacmp.v14i6.4543. [4]de Kruijf WJM,Martens RJW.Reducing patient posture variability using the predicted couch position[J].Med Dosim,2015,40(3):218-221.DOI:10.1016/j.meddos.2014.12.002. [5]McCullough KT,James JA,Cetnar AJ,et al. Site-specific tolerance tables and indexing device to improve patient setup reproducibility[J].J Appl Clin Med Phys,2015,16(3):378-388.DOI:10.1120/jacmp.v16i3.5097. [6]Hadley SW,Balter JM,Lam K L.Analysis of couch position tolerance limits to detect mistakes in patient setup[J].J Appl Clin Med Phys,2009,10(4):207-219. DOI:10.1120/jacmp.v10i4.2864. [7]De Boer HCJ,Heijmen BJM.A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload[J].Int J Radiat Oncol Biol Phys,2001,50(5):1350-1365.DOI:10.1016/S0360-3016(01)01624-8. [8]符贵山,程斌,覃仕瑞,等.放疗患者摆位误差与治疗床位置误差相关性分析[J].中华放射肿瘤学杂志,2016,25(3):266-269. DOI:10.3760/cma.j.issn.1004-4221.2016.03.015. Fu GS,Cheng B,Qin SR,et al. Correlation analysis of placement error and position error of treatment bed in patients with radiotherapy[J].Chin J Radiat Oncol,2016,25(3):266-269.DOI:10.3760/cma.j.issn.1004-4221.2016.03.015. [9]Gilbeau L,Octave-Prignot M,Loncol T,et al. Comparison of setup accuracy of three different thermoplastic masks for the treatment of brain and head and neck tumors[J].Radiother Oncol,2001,58(2):155-162. DOI:10.1016/S0167-8140(00)00280-2. [10]Wu Q,Lockman D,Wong J,et al. Effect of the first day correction on systematic setup error reduction[J].Med Phys,2007,34(5):1789-1796. DOI:10.1118/1.2727299.