Study of clinical application of a novel immobilization system in total marrow irradiation
Wang Xiaoyong,Deng Die,Gong Jian,Zhao Hongli,Wang Dajiang,Zhou Yunfeng,Xie Conghua,Liu Hui
Department of Radiochenmotherapy,Zhongnan Hospital of Wuhan University,Hubei Key Laboratory of Tumor Biological Behaviors, Hubei Cancer Clinical Study Center, Wuhan 430071,China
Objective To evaluate the clinical application value of a novel immobilization system in total marrow irradiation (TMI) with MVCT image. Methods From 2016 to 2017,a retrospective analysis of the setup errors of 22 patients receiving TMI in two groups(twelve patients were immobilized with the novel immobilization system in group 1, ten patients were immobilized with the combinatorial immobilization devices in group 2) was performed in this study on Zhongnan Hospital of Wuhan University. Two-sample t-test was used to analyze the differences of setup errors and the consistency of setup between two groups. Results In group 1,the setup errors on left-right,superior-inferior,anterior-posterior and rotation directions were (1.06±0.79),(1.34±0.66),(2.45±1.48) mm and (0.63°±0.65°) for the head and neck position,(1.58±1.13),(2.38±1.99),(2.05±1.68) mm and (0.31°±0.32°) for the chest position,(1.67±1.24),(3.88±2.20),(1.96±1.32) mm and (0.48°±0.53°) for the pelvis position,and (0.95±0.73),(1.99±1.35),(3.66±2.13) mm and (0.24°±0.31°) for the lower limb,respectively. In group 2,the setup errors were (2.59±2.58),(3.28±1.85),(3.71±2.43) mm and (1.15°±1.18°) for the head and neck position,(4.38±3.69),(5.64±3.78),(2.72±1.91) mm and (1.55°±0.86°) for the chest position,(4.14±2.97),(6.97±3.68),(2.21±2.26) mm and (1.23°±0.74°) for the pelvis position,(2.28±1.15),(5.97±3.00),(3.44±1.93) mm and (1.09°±0.94°) for the lower limb,respectively. The setup errors significantly differed between two groups on the left-right,superior-inferior and rotation directions for all positions (all P<0.05). The setup consistency significantly differed between two groups on the left-right,superior-inferior and rotation directions for the chest and pelvis positions (all P<0.05). Conclusion The novel immobilization system can significantly improve the setup accuracy and setup consistency,and enhance the precision of treatment for patients.
Wang Xiaoyong,Deng Die,Gong Jian et al. Study of clinical application of a novel immobilization system in total marrow irradiation[J]. Chinese Journal of Radiation Oncology, 2019, 28(8): 616-620.
[1] Schultheiss TE,Wong J,Liu A,et al. Image-guided total marrow and total lymphatic irradiation using helical tomotherapy[J]. Int J Radiat Oncol Biol Phys,2007,67(4):1259-1267. DOI:10.1016/j.ijrobp.2006.10.047.
[2] Bulent A,Mete Y,Gulbin OK,et al. Total marrow irradiation with rapidarc volumetric arc therapy[J]. Int J Radiat Oncol Biol Phys,2011,81(2):592-599. DOI:10.1016/j.ijrobp.2010.11.035.
[3] 胡逸民. 治疗体位及体位固定技术[A]//殷蔚伯,余子豪,徐国镇,等. 肿瘤放射治疗学[M].4版. 北京:中国协和医科大学出版社,2008:95-99.
Hu YM. Treatment of position and fixed position technology[A]//Yin WB,Yu ZH,Xu GZ,et al. Radiation Oncology[M].4th ed. Beijing:Peking union medical college press,2008:95-99.
[4] 胡丽丽,楼晓,朱峰,等. 螺旋断层治疗用于全骨髓照射的初步结果[J]. 中华放射肿瘤学杂志,2016,25(4):367-371. DOI:10.3760/cma.j.issn.1004-4221.2016.04.013.
Hu LL,Luo X,Zhu F,et al. Preliminary results of helical tomotherapy for total marrow irradiation[J]. Chin J Radiat Oncol,2016,25(4):367-371. DOI:10.3760/cma.j.issn.1004-4221.2016.04.013.
[5] 解传滨,徐寿平,俞伟,等. 容积旋转调强放射治疗技术行全骨髓照射的可行性研究[J]. 中国医学装备,2014,11(12):118-121. DOI:10.3969/J. ISSN.1672-8270.2014.12.045.
Xie CB,Xu SP,Yu W,et al. The feasibility study on total marrow irradiation with rapid arc volumetric arc therapy[J]. Chin Med Equip,2014,11(12):118-121. DOI:10.3969/J. ISSN.1672-8270.2014.12.045.
[6] 佟颖,尹勇,陈进琥,等. TomoTherapy治疗不同部位肿瘤的摆位误差[J]. 中华放射医学与防护杂志,2017,37(5):352-358. DOI:10.3760/cma.j.issn.0254-5098.2017.05.006.
Tong Y,Yin Y,Chen JH,et al. Setup errors for the tumors in different parts of body with TomoTherapy[J]. Chin J Radiol Med Prot,2017,37(5):352-358. DOI:10.3760/cma.j.issn.0254-5098.2017.05.006.
[7] 王振立,孙晓东,张耀文,等. 食管癌患者螺旋断层治疗的摆位误差分析[J]. 中华放射肿瘤学杂志,2017,26(4):429-432. DOI:10.3760/cma.j.issn.1004-4221.2017.04.013.
Wang ZL,Sun XD,Zhang YW,et al. An analysis of setup errors in helical tomotherapy for esophageal cancer patients[J]. Chin J Radiat Oncol,2017,26(4):429-432. DOI:10.3760/cma.j.issn.1004-4221.2017.04.013.
[8] Thondykandy BA,Swamidas JV,Agarwal J,et al. Setup error analysis in helical tomotherapy based image-guided radiation therapy treatments[J]. J Med Phys,2015,40(4):233-239. DOI.10.4103/0971-6203.170796.
[9] 李明,黄灿红,储开岳,等.改进型腹板在直肠癌术后放疗中的应用[J]. 肿瘤研究与临床,2014,26(6):397-399. DOI:10.3760/cma.j.issn.1006-9801.2014.06.009.
Li M,Huang CH,Chu KY,et al. Application of improved belly board in postoperative patients of rectal cancer radiotherapy[J]. Cancer Res Clin,2014,26(6):397-399. DOI:10.3760/cma.j.issn.1006-9801.2014.06.009.