Objective To introduce the therapeutic procedures of helical tomotherapy based total marrow irradiation(HT-TMI), and validate the feasibility of TMI.Methods At 1 week before bone marrow transplantation, 12 patients received TMI conditioning regimen at a prescriptive dose of 12 Gy in 3 times, once daily. Patient immobilization, CT simulation, target delineation, plan design and dosimetric verification were implemented in sequence according to the TMI procedures. The dosimetric paramaters of the target and normal tissues were analysed. The correlation between the dose verification and image-guided Results was observed.Results Compared with total body irradiation(TBI), TMI could significantly reduce the irradiation dose to normal tissues. The median dose D50 of all normal tissues except the oral cavity were lower than 6 Gy, where The D50 of lens, brain, lung and liver are (1.8±0.1) Gy, (5.7±0.2) Gy, (5.2±0.2) Gy, and (4.6±0.2) Gy, respectively. Pass rate of γ Index was larger than 95% with 3mm/3% criterian for each section. The positioningerror of head and neck was relatively lower than that of pelvis at x-axis, and was higher at z-axis.Conclusions Helical tomotherapy based TMI is a feasible and reasonable approach, which has evident dosimetric advantage.
Zhu Feng,Wu Weizhang,Chang Dongshu et al. Feasibility of helical tomotherapy intensity-modulated radiation therapy applied in total marrow irradiation[J]. Chinese Journal of Radiation Oncology, 2018, 27(1): 94-100.
[1] 胡丽丽,楼晓,朱峰,等.螺旋断层治疗用于全骨髓照射的初步结果[J].中华放射肿瘤学杂志,2016,25(4):367-371.DOI:10.3760/cma.j.issn.1004-4221.2016.04.013.Hu LL,Lou X,Zhu F,et al. Preliminary results of helical tomotherapy for total marrow irradiation[J].Chin J Radiat Oncol,2016,25(4):367-371.DOI:10.3760/cma.j.issn.1004-4221.2016.04.013
[2] 张磊.基于螺旋断层放疗系统的兆伏级CT图像分析肿瘤治疗中的摆位偏差[D].衡阳:南华大学,2016.Zhang L.A quantitative analysis of tumor site specific setup based on 1485 daily MVCT scans from helical tomo therapy[D].Hengyang:University of South China,2016
[3] 张淑慧,杨敬贤,崔广余,等.放疗定位膜制作及体位固定质量控制[J].中国医学物理学杂志,2015,32(4):554-558.DOI:10.3969/j.issn.1005-202X.2015.04.022.Zhang SH,Yang JX,Cui GY,et al. Quality control of positioning membrane production and posture fixation in radiotherapy[J].Chin J Med Phys,2015,32(4):554-558.DOI:10.3969/j.issn.1005-202X.2015.04.022
[4] Fogliata A,Cozzi L,Clivio A,et al. Preclinical assessment of volumetric modulated arc therapy for total marrow irradiation[J].Int J Radiat Oncol Biol Phys,2011,80(2):628-636.DOI:10.1016/j.ijrobp.2010.11.028
[5] 解传滨,徐寿平,徐伟,等.运用ArcCHECK实施旋转调强全骨髓照射超长靶区的计划验证[J].中国医疗器械杂志,2015,39(1):68-71.DOI:10.3969/j.issn.1671-7104.2015.01.019.Xie CB,Xu SP,Xu W,et al. Patient-specific dose verification method using ArcCHECK for total marrow irradiation with intensity modulated Arc therapy[J].Chin J Med lnstrument,2015,39(1):68-71.DOI:10.3969/j.issn.1671-7104.2015.01.019
[6] 文婷,黎静,张晋建,等.全身照射螺旋断层放疗剂量验证与分析[J].广东医学,2015,36(16):2474-2475.Wen T,Li J,Zhang JJ,et al. Dose verification and analysis of spiral sectional radiotherapy for whole body irradiation[J]. Guangdong Med,2015,36(16):2474-2475
[7] 吴伟章,朱夫海,王勇,等.旋转照射剂量测量仪(ArcCHECK)在螺旋断层放疗计划剂量验证中的应用[J].肿瘤预防与治疗,2014,27(1):20-23.DOI:10.3969/j.issn.1674-0904.2014.01.005.Wu WZ,Zhu FH,Wang Y,et al. Application of ArcCHECK in dose verification of tomotherapy planning[J].J Cancer Control Treat,2014,27(1):20-23.DOI:10.3969/j.issn.1674-0904.2014.01.005
[8] Wong JYC,Liu A,Schultheiss T,et al. Targeted total marrow irradiation using three-dimensional image-guided tomographic intensity-modulated radiation therapy:an alternative to standard total body irradiation[J].Biol Blood Marrow Transplant,2006,12(3):306-315.DOI:10.1016/j.bbmt.2005.10.026
[9] Hui SK,Kapatoes J,Fowler J,et al. Feasibility study of helical tomotherapy for total body or total marrow irradiation[J].Med Phys,2005,32(10):3214-3224.DOI:10.1118/1.2044428
[10] Schultheiss TE,Wong J,Liu A,et al. Image-guided total marrow and total lymphatic irradiation using helical tomotherapy[J].Int J Radiat Oncol Biol Phys,2007,67(4):1259-1267.DOI:10.1016/j.ijrobp.2006.10.047
[11] Aydogan B,Yeginer M,Kavak GO,et al. Total marrow irradiation with RapidArc volumetric arc therapy[J].Int J Radiat Oncol Biol Phys,2011,81(2):592-599.DOI:10.1016/j.ijrobp.2010.11.035
[12] Han CH,Schultheisss TE,Wong JYC.Dosimetric study of volumetric modulated arc therapy fields for total marrow irradiation[J].Radiother Oncol,2012,102(2):315-320.DOI:10.1016/j.radonc.2011.06.005
[13] Shueng PW,Lin SC,Chong NS,et al. Total marrow irradiation with helical tomotherapy for bone marrow transplantation of multiple myeloma:first experience in Asia[J].Technol Cancer Res Treat,2009,8(1):29-37.DOI:10.1177/153303460900800105
[14] Zeverino M,Agostinelli S,Taccini G,et al. Advances in the implementation of helical tomotherapy-based total marrow irradiation with a novel field junction technique[J].Med Dosim,2012,37(3):314-320.DOI:10.1016/j.meddos.2011.12.001
[15] Takahashi Y,Verneris MR,Dusenbery KE,et al. Peripheral dose heterogeneity due to the thread effect in total marrow irradiation with helical tomotherapy[J].Int J Radiat Oncol Biol Phys,2013,87(4):832-839.DOI:10.1016/j.ijrobp.2013.07.017
[16] Mancosu P,Navarria P,Reggiori G,et al. In-vivo dosimetry with Gafchromic films for multi-isocentric VMAT irradiation of total marrow lymph-nodes:a feasibility study[J].Radiat Oncol,2015,10:86.DOI:10.1186/s13014-015-0391-y
[17] Takahashi Y,Hui SK.Fast,simple,and informative patient-specific dose verification method for intensity modulated total marrow irradiation with helical tomotherapy[J].Radiat Oncol,2014,9:34.DOI:10.1186/1748-717X-9-34
[18] Takahashi Y,Hui SK.Impact of very long time output variation in the treatment of total marrow irradiation with helical tomotherapy[J].Radiat Oncol,2013,8:123.DOI:10.1186/1748-717X-8-123
[19] Mege JP,Sun WZ,Veres A,et al. Evaluation of MVCT imaging dose levels during helical IGRT:comparison between ion chamber,TLD,and EBT3 films[J].J Appl Clin Med Phys,2016,17(1):143-157.DOI:10.1120/jacmp.v17i1.5774