Abstract:Objective To investigate MRI localization the accuracy and correct the deviation for gamma knife treatment. Methds With 25-point-matrix tank, the deviation of MRI localization and its regularities could be identified after the comparison between the coordinates of MRI localization and the ones which have already been verified by CT within the deviation of 0.5 mm. Then the original MRI coordinates will be corrected by the acquired mean deviation and the geometric distortion of images. Afterwards the corrected coordinates will be compared with the standard ones and finally validated by exposure film. ResultsThere are no significant deviations on x-and z-axis after measurement in three hospitals, y-axis, however, bears deviation of (1.94±0.45) mm for hospital A,(-2.22±0.29) mm for hospital B,(-1.25±0.21) mm for hospital C,respectively. Furthermore there also exists geometric distortion of 1% on y-axis in hospital A. The corrected coordinates on y-axis (yc) will be calculated from the formula:Yc=(Y-M)+GD (Y0-Y)(Y:the original coordinates on y-axis, M:the mean of deviation on y-axis, GD:the geometric distortion,Y0:the coordinate on y-axis of the central point among the 25-point matrix).Once completed, the corrected coordinates of MRI localization is of no significant difference with the standard coordinates verified by CT. Even the deviation of focal spot on validation film is within 0.5 mm. Conclusions The 25-point-matrix tank in the multi-point measurement of the accuracy and the correction of deviation for gamma knife treatment is feasible to determine whether MRI can be utilized in the localization for head gamma knife treatment.
Li Shengyuan,Li Xiaoyang,Yan Zheng et al. Detection and correction of the accuracy of gamma knife treatment positioned by MRI[J]. Chinese Journal of Radiation Oncology, 2014, 23(6): 535-539.
[1] 李胜元,高立,周宣,等.伽玛刀治疗精度检测初探[J].总装备部医学学报,2012,14:209-211. [2] 李胜元,李骁杨,高立,等.自测伽玛刀治疗精度的探讨[J].中华放射肿瘤学杂志,2013,22:318-321. [3] Aoyama H, Shirato H, Nishioka T, et al. Magnetic resonance imaging system for three-dimensional conformal radiotherapy and its impact on gross tumor volume delineation of central nervous system tumors[J]. Int J Radiat,2001,50:821-827. [4] 张宜培,张雪宁,刘东,等.伽玛刀治疗磁共振影像定位精度的影响因素[J].中华放射医学与防护杂志,2012,32:383-385. [5] 中华人民共和国卫生部.医用磁共振成像(MRI)设备影像质量检测与评价规范[S].中华人民共和国卫生行业标准(WS/T 263-2006).北京:人民卫生出版社,2007. [6] 葛丽娟,林志凯,邓君,等.磁共振设备质量控制信号值与成像液硫酸铜浓度的变化关系研究[J].中华现代影像学杂志,2009,6:129-134. [7] Yu C, Apuzzo ML, Zee CS, et al. A phantom study of the geometric accuracy of computed tomographic and magnetic resonance imaging stereotactic localization with the Leksell stereotactic system[J]. Neurosurgery,2001,48:1029-1038. [8] Guo WY, Chu WC, Wu MC, et al. An evaluation of the accuracy of magnetic-resonance-guided Gamma knife surgery[J]. Stereotact Funct Neurosurq,1996,66 Suppl 1:85-92.