An investigation of precision of full six-degree target shift corrections using the ArcCHECK system
Bai Penggang, Dai Yitao, Chen Rongqin, Li Qixin, Cheng Yanming, Chen Chuanben, Fei Zhaodong, Chen Kaiqiang, Chen Jihong
Department of Radiation Oncolog (Bai PG, Dai YT, Li QX,Cheng YM,Chen CB,Fei ZD,Chen KQ, Chen JH),The Second Daytime Ward (Chen RQ) Fujian Cancer Hospital Fujian Medical University Cancer Hospital,Fuzhou 350014,China
Abstract:Objective To investigate the precision of full six-degree target shift corrections using the ArcCHECK system. Methods Fourteen patients receiving intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma (NPC) in Fujian Medical University Cancer Hospital from May to September, 2015 were selected. The first treatment setup errors were obtained using cone-beam computed tomography. The setup errors were simulated in ArcCHECK, and the full six-degree target shift corrections was used to correct the errors. The plans without and with setup errors and the plan with corrected setup errors were taken. The paired t-test was used to compare dose to agreement (DTA) and Gamma passing rates betweenthe plan without setup errors and the plan with setup errors and plan with corrected setup errors. ResultsThe DTA and Gamma passing rates were (96.76±1.57)% and (98.35±0.92)% for the plan without setup errors,(59±21.42)% and (62.86±21.63)% for the plan with setup errors, and (91.41±4.82)% and (94.11±4.33)% for the plan with corrected setup errors. There were significant differences between the plan without setup errors and the plan with setup errors and plan with corrected setup errors in DTA passing rate (t=6.64 and 5.13, both P<0.05) and Gamma passing rate (t=6.15 and 4.19, both P<0.05). Conclusions The full six-degree target shift corrections can be used in IMRT for NPC, with good Results in correcting setup errors and improving the precision for IMRT dose distribution.
Bai Penggang,Dai Yitao,Chen Rongqin et al. An investigation of precision of full six-degree target shift corrections using the ArcCHECK system[J]. Chinese Journal of Radiation Oncology, 2018, 27(2): 195-198.
[1] Yue NJ,Knisely JPS,Song HJ,et al. A method to implement full six-degree target shift corrections for rigid body in image-guided radiotherapy[J].Med Phys,2006,33(1):21-31.DOI:10.1118/1.2138009. [2] Zhang X,Shan GP,Liu JP,et al. Margin evaluation of translational and rotational set-up errors in intensity modulated radiotherapy for cervical cancer[J].Springer Plus,2016,5:153.DOI:10.1186/s40064-016-1796-2. [3] Fu WH,Yang Y,Yue NJ,et al. Dosimetric influences of rotational setup errors on head and neck carcinoma intensity-modulated radiation therapy treatments[J].Med Dosim,2013,38(2):125-132.DOI:10.1016/j.meddos.2012.09.003. [4] Bose S,Shukla H,Maltz J.Beam-centric algorithm for pretreatment patient position correction in external beam radiation therapy[J].Med Phys,2010,37(5):2004-2016.DOI:10.1118/1.3327457. [5] Hoffmans-Holtzer NA,Hoffmans D,Dahele M,et al. Roll and pitch set-up errors during volumetric modulated arc delivery:can adapting gantry and collimator angles compensate[J].Strahlenther Onkol,2015,191(3):272-280.DOI:10.1007/s00066-014-0766-2. [6] 中国鼻咽癌临床分期工作委员会.2010鼻咽癌调强放疗靶区及剂量设计指引专家共识[J].中华放射肿瘤学杂志,2011,20(4):267-269.DOI:10.3760/cma.j.issn.1004-4221.2011.04.001.Chinese Committee for Staging of Nasopharyngeal Carcinoma. The expert consensus on design guidelines of target and dose of intensity modulated radiotherapy for nasopharygneal carcinoma in 2010[J].Chin J Radiat Oncol,2011,20(4):267-269.DOI:10.3760/cma.j.issn.1004-4221.2011.04.001. [7] 张俊俊,邱小平,李奇欣,等.ArcCHECK系统在鼻咽癌容积旋转调强剂量验证中的应用[J].中国医学物理学杂志,2014,31(5):5136-5138,5168.DOI:10.3969/j.issn.1005-202X.2014.05.009.Zhang JJ,Qiu XP,Li QX,et al. Dosimetric verification of volumetric modulated arc therapy in nasopharyngeal carcinoma using ArcCHECK system[J].Chin J Med Phys,2014,31(5):5136-5138,5168.DOI:10.3969/j.issn.1005-2Z2X.2014.05.009. [8] 李成强,李光俊,冀传仙,等.ArcCHECK半导体探头特性及在容积调强弧形治疗剂量验证应用研究[J].中华放射肿瘤学杂志,2013,22(3):253-257.DOI:10.3760/cma.j.issn.1004-4221.2013.03.024.Li CQ,Li GJ,Ji CX,et al. The characteristics and clinical application of the ArcCHECK diode array for volumetric-modulated arc therapy verification[J].Chin J Radiat Oncol,2013,22(3):253-257.DOI:10.3760/cma.j.issn.1004-4221.2013.03.024. [9] Yang K,Yan D,Tyagi N.Sensitivity analysis of physics and planning SmartArc parameters for single and partial arc VMAT planning[J].J Appl Clin Med Phys,2012,13(6):3760.DOI:10.1120/jacmp.v13i6.3760.