Catalyst系统在乳腺癌患者放疗摆位应用及影响因素分析
罗焕丽, 彭海燕, 靳富, 肖鹏, 曹绍艾, 何亚男, 董文玲, 李雪敏, 杨丁懿, 王颖
400030 重庆大学附属肿瘤医院 重庆市肿瘤研究所 重庆市肿瘤医院肿瘤放疗中心
Catalyst system in patient positioning during breast cancer radiotherapy:clinical application and influencing factors
Luo Huanli, Peng Haiyan, Jin Fu, Xiao Peng, Cao Shaoai, He Yanan, Dong Wenling, Li Xuemin, Yang Dingyi, Wang Ying
Division of Physics, Department of Radiation Oncology,Chongqing University Cancer Hospital,Chongqing Cancer Institute,Chongqing Cancer Hospital,Chongqing 400030,China
全文: PDF (0 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      背景资料
摘要 目的 评价Catalyst系统在乳腺癌患者放疗摆位中的应用,并分析其与患者年龄及BMI的相关性。方法 选取 2016年5-8月间重庆市肿瘤研究所收治的 24例乳腺癌患者。每次治疗前均采用表面光学成像系统(C-Rad Catalyst)辅助摆位,每周1次kV-kV位置验证模式验证摆位。记录并观察患者的年龄、BMI以及2个系统在三维方向摆位误差。采用独立样本t检验差异,采用Pearson相关性分析。结果 C-Rad Catalyst系统与kV-kV位置验证模式的摆位误差分别为AP方向(0.22±0.17) cm与(0.18±0.13) cm (P<0.05);SI方向(0.23±0.18) cm与(0.19±0.15) cm (P>0.05);LR方向(0.28±0.28) cm与(0.20±0.15) cm (P>0.05)。kV-kV位置验证模式摆位受年龄和BMI的影响不显著(P均>0.05);而C-Rad Catalyst系统的摆位受年龄和BMI影响显著:年龄≤44岁和 45~59岁患者在AP、SI方向不同(P均<0.05),≤44岁和≥60岁患者在AP、LR方向不同(P均<0.05),45~59岁和≥60岁患者在LR方向不同(P<0.05);BMI<25 kg/m2和≥25 kg/m2患者在SI方向不同(P<0.05)。年龄≥60岁患者C-Rad Catalyst系统摆位误差在SI方向与年龄相关(r=-0.496,P<0.05);BMI<25 kg/m2患者C-Rad Catalyst系统摆位误差在AP方向与BMI相关(r=0.252,P<0.05),而BMI≥25 kg/m2患者C-Rad Catalyst系统摆位误差在SI方向与BMI相关(r=0.445,P<0.05)。结论 C-Rad Catalyst系统摆位与kV-kV位置验证模式在AP方向上差异显著,C-Rad Catalyst系统摆位易受乳腺癌患者年龄和BMI的影响。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
罗焕丽
彭海燕
靳富
肖鹏
曹绍艾
何亚男
董文玲
李雪敏
杨丁懿
王颖
关键词 乳腺肿瘤/放射疗法放射治疗计划 计算机辅助位置验证模式摆位误差    
AbstractObjective To evaluate the clinical application of Catalyst system in patient positioning during breast cancer radiotherapy, and to analyze its correlation with age and body mass index (BMI). Methods Twenty-four patients with breast cancer who were admitted to our hospital from May to August, 2016 were enrolled as subjects. For all patients, auxiliary positioning was made by the optical surface imaging system (C-Rad Catalyst) before each treatment. The kV-kV imaging was executed weekly to verify positioning. Age, BMI, and setup errors of the two systems in the anterior-posterior (AP), superior-inferior (SI), and left-right (LR) directions were recorded and analyzed by independent samples t-test and Pearson correlation analysis. Results The C-Rad Catalyst system had a significantly larger setup error in the AP direction than the kV-kV imaging (0.22±0.17 vs. 0.18±0.13 cm, P<0.05). There were no significant differences in setup errors in the SI or LR direction between the two systems (0.23±0.18 vs. 0.19±0.15 cm, P>0.05;0.28±0.28 vs. 0.20±0.15 cm, P>0.05). Age and BMI of patients had significant impacts on the C-Rad Catalyst system but the kV-kV imaging (P>0.05):there were significant differences in setup errors in the AP and SI directions between patients ≤44 years of age and those 45-59 years of age (all P<0.05);there were significant differences in setup errors in the AP and LR directions between patients ≤44 years of age and those ≥60 years of age (all P<0.05);there was a significant difference in setup error in the LR direction between patients 45-59 years of age and those ≥60 years of age (P<0.05);there was a significant difference in setup error in the SI direction between patients with BMIs of<25 and ≥25 kg/m2(P<0.05). For patients ≥60 years of age, setup error of the C-Rad Catalyst system in the SI direction was correlated with age (r=-0.496, P<0.05). For patients with BMI of<25 kg/m2, setup error of the C-Rad Catalyst system in the AP direction was correlated with BMI (r=-0.445,P<0.05). For patients with a BMI of ≥25 kg/m2, setup error of the C-Rad Catalyst system in the SI direction was correlated with BMI (r=-0.252,P<0.05). Conclusions There is significant difference in setup error in the AP direction between the C-Rad Catalyst system and the kV-kV imaging. Age and BMI have impacts on patient positioning by the C-Rad Catalyst system.
Key wordsBreast neoplasms/radiotherapy    Radiotherapy planning, computer-assisted    C-Rad Catalyst    Setup error   
收稿日期: 2017-01-22     
基金资助:国家自然科学基金(11575038);重庆市卫生计生委医学科研项目(2015MSXM229)
通讯作者: 靳富,Email:jfazj@126.com;王颖,Email:wangying_cq@hotmail.com   
引用本文:   
罗焕丽,彭海燕,靳富等. Catalyst系统在乳腺癌患者放疗摆位应用及影响因素分析[J]. 中华放射肿瘤学杂志, 2018, 27(2): 190-194.
Luo Huanli,Peng Haiyan,Jin Fu et al. Catalyst system in patient positioning during breast cancer radiotherapy:clinical application and influencing factors[J]. Chinese Journal of Radiation Oncology, 2018, 27(2): 190-194.
链接本文:  
http://journal12.magtechjournal.com/Jweb_fszlx/CN/10.3760/cma.j.issn.1004-4221.2018.02.014     或     http://journal12.magtechjournal.com/Jweb_fszlx/CN/Y2018/V27/I2/190
京公网安备11010502022167号
版权所有 © 2010《中华放射肿瘤学杂志》编辑部
地址:北京朝阳区潘学园南里17号 中国医学科学院肿瘤医院(100021)
电话:010-67700737,87788294 Email: cjron@cmaph.org
本系统由北京玛格泰克科技发展有限公司设计开发 技术支持:Support@magtech.com.cn