Effect of rotational errors on dose of positive lymph nodes in cervical cancer VMAT plan and its coping strategies
Xu Peng1, Cao Lu1, Deng Qiao1, Peng Bigeng1, Li Suping2, Li Xianfu1
1Department of Oncology, Affiliated Hospital of North Sichuan Medical College, Nanchong 637000, China; 2Department of Nuclear Medicine, Affiliated Hospital of North Sichuan Medical College, Nanchong 637000, China
AbstractObjective To evaluate the effect of rotational errors (antero-posterior) on dosimetric parameters of positive lymph nodes in the long target volumetric modulated arc therapy (VMAT) plan for advanced cervical cancer and investigate its coping strategies. Methods Clinical data of patients with cervical cancer complicated with para-aortic or inguinal lymph node metastasis admitted to Affiliated Hospital of North Sichuan Medical College were randomly selected and retrospectively analyzed. The target areas of the lymph nodes at different distances from the center of the plan were outlined according to the requirements. After designing the VMAT plan on the CT images of each case, the rotational errors (antero-posterior) were introduced by changing the parameters of the treatment couch, and the dose distribution was reconstructed by dose calculation with other parameters unchanged. Then, the external boundary of the original lymph node target was added according to d=2πr(α/360) (r is the distance from the center of the lymph node to the plan center), re-planned, and the changes of dosimetric parameters in the target area of the original lymph node were analyzed after the corresponding rotational errors were introduced. Results When the distance between the lymph node target area and the plan center was 6 cm with an error of 3°, the distance was 9 cm and 12 cm with an error of 2.5°, the distance was 15 cm with an error of 2°, and the distance was 18 cm with an error of 1.5°, the mean change of D95% was more than 5%. When the rotational errors were ≤1°, the mean change of D95% in lymph node target area was less than 5%, and when the lymph node was 18 cm away from the treatment plan center, the mean change was more than 3%, reaching 3.75%. When the rotational errors were 0.5° and the distance from the plan center was 18 cm (0.5°, 18 cm), the dose change of lymph node target was more than 5%, reaching 5.58%. At (1°, 15 cm), the V100% change reached 8.96%, and at (1°, 18 cm), the V100% change was 14.5%. The D95% and V100% parameters of the original lymph node target were changed by less than 1% after adding the external boundary of the original lymph node target and introducing corresponding rotational errors. Conclusions In the long target area radiotherapy of cervical cancer, the variation of dosimetric parameters of lymph node target was increased with the increase of rotational errors and with the increase of distance from the plan center. It is recommended to increase the efferent boundary of lymph nodes in different positions to avoid underdose by d=2πr(α/360).
Fund:Foundation of Sichuan Provincial Health Commission (19PJ037); Cooperative Scientific Research Project of Nanchong City and School (18SXHZ0398)
Corresponding Authors:
Li Xianfu, Email: lixianfu13@163.com
Cite this article:
Xu Peng,Cao Lu,Deng Qiao et al. Effect of rotational errors on dose of positive lymph nodes in cervical cancer VMAT plan and its coping strategies[J]. Chinese Journal of Radiation Oncology, 2023, 32(1): 36-41.
Xu Peng,Cao Lu,Deng Qiao et al. Effect of rotational errors on dose of positive lymph nodes in cervical cancer VMAT plan and its coping strategies[J]. Chinese Journal of Radiation Oncology, 2023, 32(1): 36-41.
[1] Bray F, Ferlay J, Soerjomataram I, et al.Global cancer statistics 2018: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries[J]. CA Cancer J Clin, 2018,68(6):394-424. DOI: 10.3322/caac.21492.
[2] 武雅琴, 韩晶晶, 朱必清, 等. 宫颈癌腹主动脉旁淋巴结转移静态调强放疗与容积旋转调强放疗的剂量学比较[J].南京医科大学学报(自然科学版),2018,38(9):1275-1279,1291. DOI: 10.7655/NYDXBNS20180920.
Wu YQ, Han JJ, Zhu BQ, et al.Dosimetric study of intensity modulated radiotherapy and volumetric-modulated arc radiotherapy for cervical cancer with para-aortic lymph node metastasis[J]. Journal of Nanjing Medicial University,2018,38(9):1275-1279,1291. DOI: 10.7655/NYDXBNS20180920.
[3] Gunderson LL, Tepper JE, Bogart JA.Clinical Radiation Oncology[M]. 3rd ed. Philadelphia, PA: Elsevier Saunders, 2012.
[4] Sapienza LG, Gomes M, Calsavara VF, et al.Does para-aortic irradiation reduce the risk of distant metastasis in advanced cervical cancer? A systematic review and meta-analysis of randomized clinical trials[J]. Gynecol Oncol, 2017,144(2):312-317. DOI: 10.1016/j.ygyno.2016.11.044.
[5] Abu-Rustum NR, Yashar CM, Bean S, et al.NCCN guidelines insights: cervical cancer, version 1.2020[J]. J Natl Compr Canc Netw, 2020,18(6):660-666. DOI: 10.6004/jnccn.2020.0027.
[6] 曹璐. 宫颈癌VMAT放疗摆位误差及利用3DVH系统分析摆位误差对剂量影响[D]. 南充:川北医学院, 2019.
Cao L.The set-up errors and impact of the errors on dose distribution in volumetric modulated arc therapy for cervical cancer based on 3DVH system[D]. Nanchong: North Sichuan Medical College, 2019.
[7] 黄叶才, 漆云翔, 任静, 等. 磁共振成像对宫颈癌小淋巴结转移的诊断价值和对放射治疗靶区设计的影响[J].肿瘤防治研究,2014,41(8):920-924. DOI: 10.3971/j.issn.1000-8578.2014.08.014.
Huang YC, Qi YX, Ren J, et al.Diagnostic value of MRI in detection of small lymph node metastasis and its effects on radiotherapy target volume design in cervical cancer patients[J]. Cancer Research on Prevention and Treatment,2014,41(8):920-924. DOI: 10.3971/j.issn.1000-8578.2014.08.014.
[8] Laursen LV, Elstrøm UV, Vestergaard A, et al.Residual rotational set-up errors after daily cone-beam CT image guided radiotherapy of locally advanced cervical cancer[J]. Radiother Oncol, 2012,105(2):220-225. DOI: 10.1016/j.radonc.2012.08.012.
[9] Wakatsuki M, Ohno T, Kato S, et al.Impact of boost irradiation on pelvic lymph node control in patients with cervical cancer[J]. J Radiat Res, 2014,55(1):139-145. DOI: 10.1093/jrr/rrt097.
[10] Liu Y, Yu J, Qian L, et al.Extended field intensity-modulated radiotherapy plus concurrent nedaplatin treatment in cervical cancer[J]. Oncol Lett, 2016,11(5):3421-3427. DOI: 10.3892/ol.2016.4416.
[11] Cihoric N, Tapia C, Krüger K, et al.IMRT with 18FDG-PET/CT based simultaneous integrated boost for treatment of nodal positive cervical cancer[J]. Radiation oncology, 2014,9(1): 83. DOI: 10.1186/1748-717X-9-83.
[12] 李志聪, 向昭雄, 游雁, 等. 旋转摆位误差对调强放疗剂量准确性的影响[J].医疗卫生装备,2016,37(4):109-111. DOI: 10.7687/J.ISSN1003-8868.2016.04.109.
Li ZC, Xiang ZX, You Y, et al.Impact of rotational set-up error in dose veracity of intensity-modulated radiotherapy[J].Chinese Medical Equipment Journal,2016,37(4):109-111. DOI: 10.7687/J.ISSN1003-8868.2016.04.109.
[13] Morice P, Castaigne D, Pautier P, et al.Interest of pelvic and paraaortic lymphadenectomy in patients with stage IB and II cervical carcinoma[J]. Gynecol Oncol, 1999,73(1):106-110. DOI: 10.1006/gyno.1998.5308.
[14] Beadle BM, Jhingran A, Yom SS, et al.Patterns of regional recurrence after definitive radiotherapy for cervical cancer[J]. Int J Radiat Oncol Biol Phys, 2010,76(5):1396-1403. DOI: 10.1016/j.ijrobp.2009.04.009.
[15] 刘礼东, 杨振, 钟美佐, 等. 肿瘤放疗中塑形垫结合热塑网膜体位固定方式的摆位误差研究[J].中国肿瘤临床,2020,47(4):198-201. DOI: 10.3969/j.issn.1000-8179. 2020.04.137.
Liu LD, Yang Z, Zhong MZ, et al.Analysis of setup errors in patients using the customized Klarity AccuCushion? With a thermoplastic fixation mask while receiving radiotherapy[J]. Chinese Journal of Clinical Oncology,2020,47(4):198-201. DOI: 10.3969/j.issn.1000-8179.2020.04.137.
[16] Noufal MP, Abdullah KK, Niyas P, et al.Analysis of influence of errors in angular settings of couch and collimator on the dosimetric and radiobiological parameters in VMAT plans[J]. J Med Imaging Radiat Sci, 2017,48(2):166-177. DOI: 10.1016/j.jmir.2016.10.012.
[17] 王占宇, 谭军文, 龙雨松, 等. 基于剂量重建分析摆位误差对鼻腔NK/T细胞淋巴瘤剂量学影响[J].中华放射肿瘤学杂志,2019,28(11):849-853. DOI: 10.3760/cma.j.issn.1004-4221.2019.11.011.
Wang ZY, Tan JW, Long YS, et al.Dosimetric effect of set-up errors on nasal NK/T cell lymphoma based on dose reconstruction[J]. Chinese Journal of Radiation Oncology, 2019,28(11): 849-853. DOI: 10.3760/cma.j.issn.1004-4221.2019.11.011.
[18] 李陆军, 朱泓政, 游雁, 等. 电子射野影像系统引导鼻咽癌调强放射治疗中不同图像配准范围的对比研究[J].中华放射医学与防护杂志,2015,35(5):363-365. DOI: 10.3760/cma.j.issn.0254-5098.2015.05.011.
Li LJ, Zhu HZ, You Y, et al.Comparative study of different image registration ranges in intensity modulated radiotherapy for nasopharyngeal carcinoma guided by electronic field imaging system[J]. Chinese Journal of Radiological Medicine and Protection,2015,35(5):363-365. DOI: 10.3760/cma.j.issn.0254-5098.2015.05.011.
[19] Klein EE, Hanley J, Bayouth J, et al.Task group 142 report: quality assurance of medical accelerators[J]. Med Phys, 2009,36(9):4197-4212. DOI: 10.1118/1.3190392.
[20] Sethi R, Mayadev J, Sethi S, et al.Patterns of recurrence in node-positive cervical cancer patients treated with contemporary chemoradiation and dose escalation: a multi-institutional study[J]. Pract Radiat Oncol, 2019,9(2): e180-e186. DOI: 10.1016/j.prro.2018.08.012.