The effect of age on skin elasticity and setup error in optical surface image-guided radiotherapy
He Yang, Lei Jinyan, Peng Haiyan, Luo Huanli, Mao Kaijin, Wang Dengyan, Peng Lifeng, Zhu Yiyu, Min Qinghong, Wang Xuechun, Tang Huan, Xiao Peng, Xiao Qiang, Wang Ying, Jin Fu
Department of Radiation Oncology, Chongqing Cancer University Cancer Hospital/Chongqing Cancer Institute/Chongqing Cancer Hospital, Chongqing 400030, China
AbstractObjective To investigate the correlation between skin elasticity and setup error in optical surface image-guided radiotherapy. Methods The skin elasticity (R7) data of the head, chest and abdomen were extracted and analyzed its correlation with age by systematic literature review. Fifty-four patients diagnosed with nasopharyngeal carcinoma, breast cancer and cervical cancer were recruited in this study. Firstly,the patients were positioned based on the room laser and markers. Subsequently, the patient position was verified by the Varian On-Board Imager, and then C-Rad Catalyst was adopted to obtain surface images in two states (mask or non-mask) as reference images. In the subsequent fraction treatment,after initial positioning,the local calibration was performed by Catalyst,and setup errors in three directions were recorded. Meanwhile,the patient setup was verified by CBCT twice a week. The Pearson correlation analysis was performed to analyze the correlation between setup error and age. Results The skin elasticity was negatively correlated with aging (P<0.01). The correlation coefficient between random error and age in head-and-neck cancer were 0.645,0.624 and 0.866 in the AP,SI and LR directions (all P<0.05) for male patients without mask,respectively. The system error was significantly correlated with age in the LR direction (P<0.05) for male patients,and in the AP direction (P<0.05) for female patients with head-and-neck cancer without mask. The setup error had a significant correlation with skin elasticity in male patients with head-and-neck cancer,and the sequence of absolute value of correlation coefficient was LR>SI>AP. Conclusion In optical surface-guided radiotherapy of head and neck cancer,skin elasticity may be a significant index for assessing the setup errors in male patients.
He Yang,Lei Jinyan,Peng Haiyan et al. The effect of age on skin elasticity and setup error in optical surface image-guided radiotherapy[J]. Chinese Journal of Radiation Oncology, 2020, 29(1): 47-51.
He Yang,Lei Jinyan,Peng Haiyan et al. The effect of age on skin elasticity and setup error in optical surface image-guided radiotherapy[J]. Chinese Journal of Radiation Oncology, 2020, 29(1): 47-51.
[1]胡健,徐利明,姚颐,等. KV-KV、KV-MV和CBCT在头颈部调强放疗位置验证中的对比研究[J]. 中国医学物理学杂志,2010,27(4):1988-1992. DOI:10.3969/j.issn.1005-202X.2010.04.010.
Hu J Xu LM,Yao Y,et al. The Comparation of 2D kilovoltage with kilovoltage-megavoltage radiographs and cone-beam computed tomography in position verification of head and neck intensity modulate radiotherapy[J]. Chin J Med Phys,2010,27(4):1988-1992. DOI:10.3969/j.issn.1005-202X.2010.04.010.
[2]郎锦义,王培,吴大可,等.2015年中国大陆放疗基本情况调查研究[J]. 中华放射肿瘤学杂志,2016,25(6):541-545. DOI:10.3760/cma.j.issn.1001-4221.2016.06.001.
Lang JY,Wang P,Wu DK,et al. An investigation of the basic situation of radiotherapy in mainland China in 2015[J]. Chin J Radiat Oncol,2016,25(6):541-545. DOI:10.3760/cma.j.issn.1001-4221.2016.06.001.
[3]Stieler F,Wenz F,Shi M,et al. A novel surface imaging system for patient positioning and surveillance during radiotherapy[J]. Strahlenther Onkol,2013,189(11):938-944. DOI:10.1007/s00066-013-0441-z.
[4]朱南南,张骏. 表面粗糙度激光散射检测的多波长光纤传感器[J]红外与激光工程,2016,45(5):217-223. DOI:10.3788/IRLA201645.0522003.
[5]Ma L,Tan Y,Zheng S,et al. Correlation study between image features and mechanical properties of Han Chinese facial skin[J]. Int J Cosmet Sci,2017,39(1):93-100. DOI:10.1111/ics.12356.
[6]Oikarinen A. The aging of skin:chronoaging versus photoaging[J]. Photodermatol Photoimmunol Photomed,1990,7(1):3-4. DOI:10.1016/0190-9622(90)70053-K
[7]Dobrev H. Cutometer® [M]. Berlin:Springer Heidelberg,2014:315-338.
[8]Ryu HS,Joo YH,Sun OK,et al. Influence of age and regional differences on skin elasticity as measured by the Cutometer[J]. Skin Res Technol,2008,14(3):354-358. DOI:10.1111/j.1600-0846.2008.00302.x.
[9]Ahn S,Kim S,Lee H,et al. Correlation between a Cutometer® and quantitative evaluation using Moire topography in age-related skin elasticity[J]. Skin Res Technol,2007,13(3):280–284. DOI:10.1111/j.1600-0846.2007.00224.x.
[10]Nelec B,Forget N J,Hurtubise T,et al. Skin characteristics:normative data for elasticity,erythema,melanin,and thickness at 16 different anatomical locations[J]. Skin Res Technol,2016,22(3):263-275. DOI:10.1111/srt.12256.
[11]Prisciandaro JI,Frechette CM,Herman MG,et al. A methodology to determine margins by EPID measurements of patient setup variation and motion as applied to immobilization devices[J]. Med Phys,2004,31(11):2978-2988. DOI:10.1118/1.1800712.
[12]Draaijers LJ,Botman YAM,Tempelman FRH,et al. Skin elasticity meter or subjective evaluation in scars:a reliability assessment[J]. Burns,2004,30(2):109-114. DOI:10.1016/j.burns.2003.09.003.
[13]Ohshima H,Kinoshita S,Oyobikawa M,et al. Use of Cutometer area parameters in evaluating age-related changes in the skin elasticity of the cheek[J]. Skin Res Technol,2013,19(1):238-242. DOI:10.1111/j.1600-0846.2012.00634.x.
[14]潘启勇,丁秋娥,姜仁伟,等. 盆腔肿瘤常规放疗摆位误差分析[J]. 中国实用医药,2010,5(1):47-48. DOI:10.3969/j.issn.1673-7555.2010.24.028.
[15]丁伟杰,孙菁,陈一兴,等. 肝肿瘤螺旋断层放疗中患者年龄对摆位误差的影响[J]. 肿瘤,2016,36(7):810-813. DOI:10.3781/j.issn.1000-7431.2016.77.273.