Small field output factor measurement and correction method based on IAEA report No.483
Li Minghui, Ma Pan, Tian Yuan, Miao Junjie, Men Kuo, Zhang Ke, Niu Chuanmeng, Dai Jianrong
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences ( CAMS ) and Peking Union Medical College ( PUMC),Beijing 100021,China
AbstractObjective The IAEA report No. 483 describes the latest method of small field dosimetry. The field output factor measurement and correction methods are used to improve the accuracy and consistency of the measurement results for different type detectors.Methods The field output factors from 0.6 cm×0.6 cm to 10 cm×10 cm were measured using IBA′s CC13 ionization chamber, CC01 ionization chamber, PFD semiconductor detector, EFD semiconductor detector and Razor semiconductor detector, respectively. The field output correction factors were used to correct the measurement result.Results Compared with the corrected data, the results of ionization chamber are mainly affected by the volume averaging and the fluence perturbation effect, lead to the measurement result which is 4.70% lower at 0.6 cm×0.6 cm;The results of Shielded semiconductors are mainly affected by fluence perturbation effect, lead to the measurement result which is 4.80% higher at 0.6 cm×0.6 cm. The results of unshielded semiconductors are mainly affected by energy response and fluence perturbation effect, resulting in lower measurement results at the field size>0.8 cm×0.8 cm, 2.10% lower at field size of 1.5 cm×1.5 cm, higher measurement results at field size<0.8 cm×0.8 cm and 1.1% higher at field size of 0.6 cm×0.6 cm. Before the correction, the measurement results from different types of detectors are quite different, average standard deviation is 0.016 6. After the correction, the difference among the detectors is significantly reduced, average standard deviation is 0.006 6.Conclusions For detectors such as ionization chambers and semiconductors, the field output correction factors can be used to correct the output factors of the small field to improve the accuracy and consistency of the measurement results.
Fund:National Key Projects of Research and Development of China (2016YFC0904600) ;Chinese Academy of Medical Science (CAMS) Innovation Fund for Medical Sciences (CIFMS, 2016-I2M-1-001)
Cite this article:
Li Minghui,Ma Pan,Tian Yuan et al. Small field output factor measurement and correction method based on IAEA report No.483[J]. Chinese Journal of Radiation Oncology, 2019, 28(6): 452-456.
Li Minghui,Ma Pan,Tian Yuan et al. Small field output factor measurement and correction method based on IAEA report No.483[J]. Chinese Journal of Radiation Oncology, 2019, 28(6): 452-456.
[1] Das IJ,Ding GX,AhnesjöA.Small fields:nonequilibrium radiation dosimetry[J].Med Phys,2008,35(1):206-215.DOI:10.1118/1.2815356. [2] Richard S. Lawson, Duncan White,et al. An audit of half-count myocardial perfusion imaging using resolution recovery software[J].Nucl Med Commun,2014,35(5):511-521.DOI:10.1097/mnm.0000000000000078. [3] Laub WU,Wong T.The volume effect of detectors in the dosimetry of small fields used in IMRT[J].Med Phys,2003,30(3):341-347.DOI:10.1118/1.1544678. [4] Eklund K,AhnesjöA.Modeling silicon diode energy response factors for use in therapeutic photon beams[J].Phys Med Biol,2009,54(20):6135-6150.DOI:10.1088/0031-9155/54/20/007. [5] IAEA. Dosimetry of small static fields used in external beam radiotherapy:an IAEA-AAPM international code of practice for reference and relative dose determination[DB/OL][2018-03-25].https://www.iaea.org/publications/11075. [6] R. Alfonso, P. Andreo, et al. A new formalism for reference dosimetry of small and nonstandard fields[J].Med Phys,2008,35(11):5179-5186.DOI:10.1118/1.3005481. [7] Benmakhlouf H,Sempau J,Andreo P.Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams:a PENELOPE Monte Carlo study[J].Med Phys,2014,41(4):041711.DOI:10.1118/1.4868695. [8] C. Bassinet, C. Huet,et al. Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife andlinear accelerators equipped with microMLC and circular cones[J]. Med Phys,2013,40(7):071725.DOI:10.1118/1.4811139. [9] Seuntjens J.On the detector response and the reconstruction of the source intensity distribution in small photon fields[D].Quebec:McGill university,2016. [10] Czarnecki D,Zink K.Monte Carlo calculated correction factors for diodes and ion chambers in small photon fields[J].Phys Med Biol,2013,58(8):2431.DOI:10.1088/0031-9155/58/8/2431. [11] Tanny S,Sperling N,Parsai E.Correction factor measurements for multiple detectors used in small field dosimetry on the Varian Edge radiosurgery system[J].Med Phys,2015,42(9):5370-5376.DOI:10.1118/1.4928602.