Abstract Through the analysis of 3 maintenance of Bmag interlocking fault of Varian Clinac IX Linear Accelerator, it is found that the full wave rectifier diode breakdown in the bend magnet power supply, the improper setting of the bend magnet interlock, and the failure of the modulator CB3 air switch can cause the bmag interlocking. Through the replacement of accessories, adjustment parameters, etc., so that the fault is resolved.
Corresponding Authors:
Liu Xuhong, Email:3252232134@qq.com
Cite this article:
Chen Xiao,Liu Xuhong. BMAG interlock troubleshooting of Varian clinac IX linear accelerator[J]. Chinese Journal of Radiation Oncology, 2019, 28(5): 398-399.
Chen Xiao,Liu Xuhong. BMAG interlock troubleshooting of Varian clinac IX linear accelerator[J]. Chinese Journal of Radiation Oncology, 2019, 28(5): 398-399.
[1]王卫,黄英,汪松林,等.VARIAN 2300C/D直线加速器偏转磁铁的工作原理及BMAG 连锁的维修[J].医疗设备信息,2003,18(3):62-63. Wang W,Huang Y,Wang SL,et al. Working principle of deflection magnet of VARIAN 2300C/D linear accelerator and maintenance of BMAG interlock[J].Med Dev Informat.2003,18(3):62-63. [2]潘继鹏.瓦里安23EX医用直线加速器故障维修 2例[J].中国医疗设备,2016,31(4):165-166.DOI:10.3969/j.issn.1674-1633.2016.04.051. Pan JP.Two cases of fault maintenance of Varian 23EX medical linear accelerator[J].Chin. Med Equip,2016,31(4):165-166.DOI:10.3969/j.issn.1674-1633.2016.04.051. [3]曾勇,苏善宁,唐志全,等.VARIAN 高能加速器束流偏转系统[J]. 中国医疗器械信息,2017,23(07):77-80.DOI:10.15971/j.c-nki.cmdi.2017.07.022. Zeng Y,Su SN,Tang ZQ,et al. VARIAN high energy accelerator beam deflection system[J].Chin Med Dev informat.2017,23(07):77-80.DOI:10.15971/j.cnki.cmdi.2017.07.022.