Evaluation of the feasibility of log file-based three-dimensional independent dose verification system in quality assurance of intensity-modulated radiation therapy
Zhao Hanyi, Chang Sheng, Wang Dajiang, Zhang Yuemei, Bai Long, Li Guangjun
Radiation Physics Center,Cancer Center,WestChina Hospital,Sichuan University,Chengdu 610041,China (Zhao HY Present unit:Department of Oncology,Tongji Hospital,Tongji Medical College,Huazhong University of Science and Technology,Wuhan 430030,China)
AbstractObjective To assess the feasibility of the log file-based three-dimensional independent dose verification system for the quality assurance of clinical radiotherapy. Methods The statistical values of the percentage depth dose,off-axis curves and output factor calculated by the Mobius system were statistically compared with the measured data by three-dimensional water tank. The three-dimensional independent dose verification in clinical radiotherapy plan and the acceleratr log file-based three-dimensional dose verification during the treatment were performed in 17 patients with nasopharyngeal cancer. The accuracy of dose calculation and reconstruction of Mobius system was assessed. A statistical analysis was performed on the intra-fractionalγpass rate (3%/3 mm) for each patient to evaluate the stability of intra-fractional radiotherapy. Results The percentage depth dose,off-axis curve and output factor statistically calculated by the Mobius system matched well with the data measured by three-dimensional water tank. The dose-volume histogram (DVH) parameters between the target area and organ at risk during clinical radiotherapy plan were statistically compared in 17 patients with nasopharyngeal cancer. The maximum deviation was-2.16% for the three-dimensional independent dose verification in the clinical radiotherapy plan,and 0.18 Gy for the accelerator log file-based three-dimensional dose verification. The averageγpass rate for 17 nasopharyngeal cancer patients was 99.26%,and the maximum deviation of intra-fractional radiotherapy was below 0.5%. Conclusions The function of dose reconstruction and independent calculation of the Mobius system yeilds the same accuracy with the treatment planning system,which can quickly perform three-dimensional independent dose verification in the clinical radiotherapy plan and accelerator log file-based three-dimensional dose verification throughout the treatment,thereby guarantting and providing the safe and reliable technical support for clinical treatment.
Fund:National Natural Science Foundation of China (81472807);Sichuan Provincial Health and Family Planning Commission Research Project (16PJ321);Undergraduate Training Program for Innovation and Entrepreneurship of Sichuan University (201610610721)
Corresponding Authors:
Li Guangjun,Email: gjnick829@sina.com
Cite this article:
Zhao Hanyi,Chang Sheng,Wang Dajiang et al. Evaluation of the feasibility of log file-based three-dimensional independent dose verification system in quality assurance of intensity-modulated radiation therapy[J]. Chinese Journal of Radiation Oncology, 2018, 27(10): 920-924.
Zhao Hanyi,Chang Sheng,Wang Dajiang et al. Evaluation of the feasibility of log file-based three-dimensional independent dose verification system in quality assurance of intensity-modulated radiation therapy[J]. Chinese Journal of Radiation Oncology, 2018, 27(10): 920-924.
[1] 胡逸民.肿瘤放射物理学[M].北京:原子能出版社,1999:36-37. Hu YM.Radiophysics of tumor[M].Beijing:Atomic energy press,1999:36-37. [2] Xing L,Lin ZX,Donaldson SS,et al. Dosimetric effects of patient displacement and collimator and gantry angle misalignment on intensity modulated radiation therapy[J].Radiat Oncol,2000,56(1):97-108.DOI:10.1016/S0167-8140(00)00192-4. [3] Georg D,Stock M,Kroupa B,et al. Patient-specific IMRT verification using independent fluence-based dose calculation software:experimental benchmarking and initial clinical experience[J].Phys Med Biol,2007,52(16):4981-4992.DOI:10.1088/0031-9155/52/16/018. [4] Nelms BE,Zhen H,Tomé WA.Per-beam,planar IMRT QA passingrates do not predict clinically relevant patient dose errors[J].Med Phys,2011,38(2):1037-44.DOI:10.1118/1.3544657. [5] Agnew CE,King RB,Hounsell AR,et al. Implementation of phantom-less IMRT delivery verification using Varian DynaLog files and R/V output[J].Phys Med Biol,2012,57(21):6761.DOI:10.1088/0031-9155/57/21/6761. [6] 马阳光,张可,胡志辉,等.利用加速器日志文件验证容积旋转调强放疗计划的可行性[J].中华放射医学与防护杂志,2012,32(3):285.DOI:10.3760/cma.j.issn.0254-5098.2012.03.016. Ma YG,Zhang K,Hu ZH,et al. Use accelerator log files to verify the feasibility of volumetric rotational intensity modulated radiotherapy[J].Chin J Radiol Med Protect,2012,32(3):285.DOI:10.3760/cma.j.issn.0254-5098.2012.03.016. [7] Nelson C,Mason B,Robinson RC,et al. Commissioning results of an automated treatment planning verification system[J].J Appl Clin Med Phys,2014,15(5):57–65.DOI:10.1120/jacmp.v15i5.4838. [8] Jones D.ICRU Report 50:prescribing,recording and reporting photon beam therapy[J].Med Phys,1994,21(6):833-834.DOI:10.1118/1.597396. [9] Gloi AM,Buchanan RE,Zuge C L,et al. RapidArc quality assurance through MapCHECK[J].J Appl Clin Med Phys,2011,12(2):3251.DOI:10.1120/jacmp.v12i2.3251. [10] Létourneau D,Publicover J,Kozelka J,et al. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy[J]. Med Phys, 2009,36(5):1813-1821.DOI:10.1118/1.3117563. [11] Yang Y,Xing L,Li JG,et al. Independent dosimetric calculation with inclusion of head scatter and MLC transmission for IMRT[J].Med Phys,2003,30(11):2937-2947.DOI:10.1118/1.1617391. [12] Howlett SJ.Enhanced dynamic wedge and independent monitor unit verification[J].Aust Phys Eng Sci Med,2005,28(1):26-36.DOI:10.1007/BF03178861. [13] Xing L,Chen Y,Luxton G,et al. Monitor unit calculation for an intensity modulated photon field by a simple scatter-summation algorithm[J].Phys Med Biol,2000,45(3):N1-N7.DOI:10.1088/0031-9155/45/3/401. [14] Li JG,Dempsey JF,Ding L,et al. Validation of dynamic MLC-controller log files using a two-dimensional diode array.[J].Med Phys,2003,30(5):799-805. [15] Howlett SJ.Enhanced dynamic wedge and independent monitor unit verification[J].Aust Phys Eng Sci Med,2005,28(1):26-36.DOI:10.1007/BF03178861. [16] Linthout N,Verellen D,Van AS,et al. A simple theoretical verification of monitor unit calculation for intensity modulated beams using dynamic mini-multileaf collimation[J].Radiother Oncol,2004,71(2):235-241.DOI:10.1016/j.radonc.2004.02.014.