A quantitative evaluation of quality control image for on-board imaging system of medical linear accelerator
Zhuang Yongdong, Wang Bin, Zhu Jinhan, Liu Boji, Liu Xiaowei, Chen Lixin
School of Physics,Sun Yat-Sen University,Guangzhou 510275,China (Zhuang YD,Liu XW); Department of Radiation Ontology,SUN Yat-sen University Cancer Center,State Key Laboratory of Oncology in Southern China,Guangzhou 510060,China (Wang B,Zhu JH,Liu BJ,Chen LX)
AbstractObjective To establish a quantitative evaluation of quality control image for the on-board imaging system of medical linear accelerator. Methods An MV planar image of electronic portal imaging device (EPID) is acquired by both Elekta iViewGT and Varian aS1000, and a kV planar image and cone-beam computed tomography (CBCT) images of CBCT are acquired by both Elekta X-ray volume imaging (XVI) and Varian On-board Imager (OBI). Phantoms used here included Las Vegas, TOR18FG, and Catphan504. A series of image quality indicators were evaluated by analyzing the images mentioned above using a quantitative method. Results A quantitative value was calculated to represent the contrast resolution of EPID. A modulation transfer function (MTF) to describe spatial resolution and a quantitative value representing contrast resolution were calculated for the kV planar image. As for the CBCT system, a series of quantitative Results of noise, uniformity, CT value accuracy, and contrast resolution and a MTF were calculated to represent the performance of CBCT system. Conclusions Based on common phantoms, a complete set of quantitative Methods to evaluate the image quality of EPID and CBCT has been developed, which could provide a very good reference for the establishment of quality control system for image-guided radiotherapy.
Fund:Natural Science Foundation of Guangdong Province (2014A030310188);Natural Science Foundation of Guangdong Province (2016A030313276);Guangzhou Science and Technology Program (20160701168)
Corresponding Authors:
Liu Xiao-wei,Email:stslxw@mail.sysu.edu.cn;Chen Lixin,Email:chenlx@sysucc.org.cn
Cite this article:
Zhuang Yongdong,Wang Bin,Zhu Jinhan et al. A quantitative evaluation of quality control image for on-board imaging system of medical linear accelerator[J]. Chinese Journal of Radiation Oncology, 2017, 26(4): 442-447.
Zhuang Yongdong,Wang Bin,Zhu Jinhan et al. A quantitative evaluation of quality control image for on-board imaging system of medical linear accelerator[J]. Chinese Journal of Radiation Oncology, 2017, 26(4): 442-447.
[1] Pahn G,Skornitzke S,Schlemmer HP,et al. Toward standardized quantitative image quality (IQ) assessment in computed tomography (CT):a comprehensive framework for automated and comparative IQ analysis based on ICRU Report 87[J].Phys Med,2015,32(1):104-115.DOI:10.1016/j.ejmp.2015.09.017 [2] 王运来,徐寿平,戴相昆.电子射野影像系统QA与控制方法的探讨[J].中华放射肿瘤学杂志,2007,16(2):97-98.DOI:10.3760/j.issn:1004-4221.2007.02.023. Wang YL,Xu SP,Dai XK.Discussion on quality assurance and control method of electronic portal imaging system[J].Chin J Radiat Oncol,2007,16(2):97-98.DOI:10.3760/j.issn:1004-4221.2007.02.023 [3] 尹勇,袁双虎,刘同海,等.加速器附加锥形束CT图像质量评价[J].中华放射肿瘤学杂志,2007,16(3):225-228.DOI:10.3760/j.issn:1004-4221.2007.03.014. Yin Y,Yuan SH,Liu TH,et al. Evaluation of Varian kV X-ray cone beam CT image[J].Chin J Radiat Oncol,2007,16(3):225-228.DOI:10.3760/j.issn:1004-4221.2007.03.014 [4] 王运来,戴相昆,鞠忠建,等.加速器机载千伏级锥形束CT的影像质量分析[J].中华现代影像学杂志,2008,5(4):327-330. Wang YL,Dai XK,Ju ZJ,et al. Image quality evaluation for LINAC2 integrated kV cone beam CT[J].J Chin Mod Imaging,2008,5(4):327-330 [5] 王烝.医用直线加速器电子射野影像系统QA方法的研究[J].中国辐射卫生,2011,20(1):71-74. Wang Z.Research on quality assurance of amorphous silicon electronic portal imaging decice (a-Si EP ID)[J].Chin J Radiol Health,2011,20(1):71-74 [6] 柳先锋,刘翔宇,何亚男,等.Varian On-Board Imager (OBI)常规测量方法及QA[J].中国医学物理学杂志,2011,28(1):2344-2347,2394.DOI:10.3969/j.issn.1005-202X.2011.01.005. Liu XF,Liu XY,He YN,et al. Routine measurement method and quality assurance procedure on varian on-board imager[J].Chin J Med Phys,2011,28(1):2344-2347,2394.DOI:10.3969/j.issn.1005-202X.2011.01.005 [7] 汪隽琦,徐志勇,胡伟刚,等.千伏级锥形束CT图像质量和稳定性及三维影像体积重建精度研究[J].中华放射肿瘤学杂志,2012,21(2):176-180.DOI:10.3760/cma.j.issn.1004-4221.2012.02.024. Wang JQ,Xu ZY,Hu WG,et al. Quantitative analysis of image quality,stability and volume precision in kilovoltage cone beam CT[J].Chin J Radiat Oncol,2012,21(2):176-180.DOI:10.3760/cma.j.issn.1004-4221.2012.02.024 [8] 孙小喆,孟慧鹏,毛京群,等.OBI影像引导放射治疗系统的QA (QA)及维护保养[J].中国医学物理学杂志,2013,30(6):4527-4529.DOI:10.3969/j.issn.1005-202X.2013.06.015. Sun XZ,Meng HP,Mao JQ,et al. OBI image guided radiation therapy system of quality assurance (QA) and maintenance[J].Chin J Med Phys,2013,30(6):4527-4529.DOI:10.3969/j.issn.1005-202X.2013.06.015 [9] International Commission on Radiation Units and Measurements. ICRU Report No.87:radiation dose and image-quality assessment in computed tomography[J].J ICRU,2012,12(1):1-149.DOI:10.1093/jicru/ndt007. [10] Dong L,Boyer AL.An objective method for evaluating electronic portal imaging devices[J].Med Phys,1994,21(6):755-760.DOI:10.1118/1.597335. [11] Coltman JW.The specification of imaging properties by response to a sine wave input[J].J Opt Soc Am,1954,44(6):468-471.DOI:10.1364/JOSA.44.000468. [12] Droege RT,Morin RL.A practical method to measure the MTF of CT scanners[J].Med Phys,1982,9(5):758-760.DOI:10.1118/1.595124. [13] International Electrotechnical Commission (IEC).IEC 61223-3-5 Evaluation and routine testing in medical imaging departments-Part 3-5:acceptance tests–imaging performance of computed tomography X-ray equipment[S].Geneva:IEC,2004. [14] 康德华,邓小武,黄劭敏.电子射野影像系统(EPIDs)图像质量控制方法的探索[J].中国医学物理学杂志,2009,26(1):955-957. Kang DH,Deng XW,Huang SM.The quality control test for digital image of electronic portal imaging devices (EPIDs)[J].Chin J Med Phys,2009,26(1):955-957.