AbstractObjective To study the impact of setup error caused by computed tomography (CT) images with different resolutions in the Sentinel system on clinical treatment. Methods A phantom was scanned by large-aperture positioning CT with two different resolutions (CT1:0.5 mm×0.5 mm×1.0 mm, FOV 256 mm, Matrix 512, thickness 1 mm;CT3:1.0 mm×1.0 mm×3.0 mm, FOV 500 mm, Matrix 512, thickness 3 mm). The CT images were transferred to the planning system. The radiation fields were designed and transferred to MOSAIQ and Sentinel systems. Ten fixed setup errors were applied to a six degree of freedom couch. The Sentinel system was used to position the two groups of CT images and generate the setup errors. The comparison of two datasets was made by paired t-test. Cone-beam CT was used for independent verification. Results The setup errors in x-, y-, and z-directions were significantly smaller on CT1 than on CT3(0.19±0.11 vs. 0.33±0.16 mm, P=0.061;0.59±0.79 vs. 1.07±1.09 mm, P=0.008;0.67±0.75 vs. 1.16±1.30 mm, P=0.043). There were no significant differences in rotational errors in x-, y-, or z-directions between the two datasets (P=0.494;P=0.182;P=0.298). Conclusions The Sentinel system has a higher setup accuracy in the 0.5 mm×0.5 mm×1.0 mm mode than in the 1.0 mm×1.0 mm×3.0 mm mode. However, the later mode is still an acceptable choice in clinical treatment.
Fund:Health Department Foundation of Sichuan Province (110263)
Corresponding Authors:
Wang Pei. Email:dengwangpei@163.com
E-mail: dengwangpei@163.com
Cite this article:
Li Jie,Chen Chuandong,Kang Shengwei et al. Computed tomography images with different resolutions in Sentinel system:a setup error analysis[J]. Chinese Journal of Radiation Oncology, 2016, 25(8): 851-854.
Li Jie,Chen Chuandong,Kang Shengwei et al. Computed tomography images with different resolutions in Sentinel system:a setup error analysis[J]. Chinese Journal of Radiation Oncology, 2016, 25(8): 851-854.
[1] Schaly B,Bauman GS,Song W,et al. Dosimetric impact of image-guided 3D conformal radiation therapy of prostate cancer[J].Phys Med Biol,2005,50(13):3083-3101.DOI:10.1088/0031-9155/50/13/008 [2] Sykes JR,Amer A,Czajka J,et al. A feasibility study for image guided radiotherapy using low dose,high speed,cone beam X-ray volumetric imaging[J].Radiother Oncol,2005,77(1):45-52.DOI:10.1016/j.radonc.2005.05.005 [3] Walter C,Boda-Heggemann J,Wertz H,et al. Phantom and in-vivo measurements of dose exposure by image-guided radiotherapy (IGRT):MV portal images vs.kV portal images vs.cone-beam CT[J].Radiother Oncol,2007,85(3):418-423.DOI:10.1016/j.radonc.2007.10.014 [4] Marrazzo L,Simontacchi G,Agresti B,et al. Evaluation of a surface acquisition system for checking patient positioning[J].Radiother Oncol,2009,92(S1):S73.DOI:10.1016/S0167-8140(12)72782-2 [5] Pallotta S,Marrazzo L,Simontacchi G,et al. Laser/camera based system for patient setup verification:a study on phantom and patients[J].Radiother Oncol,2010,96:S529 [6] Pallotta S,Simontacchi G,Marrazzo L,et al. Accuracy of a 3D laser/camera surface imaging system for setup verification of the pelvic and thoracic regions in radiotherapy treatments[J].Med Phys,2013.40:011710.DOI:10.1118/1.4769428 [7] Krengli M,Galano S,Mones E,et al. Reproducibility of patient setup by surface image registration system in conformal radiotherapy of prostate cancer[J].Radiat Oncol,2009,4:9.DOI:10.1186/1748-717X-4-9 [8] 顾文栋,李奇林,高旻,等.002LFC调强放疗模体评价SentinelTM监控与摆位验证系统的精确性研究[J].中华放射肿瘤学杂志,2013,22(6):493-496.DOI:10.3760/cma.j.issn.1004-4221.2013.06.019 Gu WD,Li QL,Gao M,et al. A 002LFC intensity-modulated radiotherapy (IMRT) phantom evalution of SentinelTM system for patient set-up verification in radiotherapy[J].Chin J Radiat Oncol,2013,22(6):493-496.DOI:10.3760/cma.j.issn.1004-4221.2013.06.019 [9] Pallotta S,Marrazzo L,Ceroti M,et al. A phantom evaluation of SentinelTM.a commercial laser/camera surface imaging system for patient setup verification in radiotherapy[J].Med Phys,2012,39(2):706-712.DOI:10.1118/1.3675973 [10] 曾自力,刘兵,王勇兵,等.不同扫描与观察条件下三维适形放疗计划系统重建CT图像定义体积变化的研究[J].中华放射肿瘤学杂志,2010,19(4):351-354.DOI:10.3760/cma.j.issn.1004-4221.20l0.04.024 Zeng ZL,Liu B,Wang YB,et al. Impact of scanning and observing conditions on definitive volume of rebuilt CT images in three dimensional conformal radiotherapy planning system[J].Chin J Radiat Oncol,2010,19(4):351-354.DOI:10.3760/cma.j.issn.1004-4221.20l0.04.024