Abstract Objective To investigate the CT number linearity of the cone-beam CT (CBCT) images at the different spatial locations in the scanning area. Methods The Catphan504 phantom at the different locations are scanned repeatedly using the CBCT on the Varian 23EX linear accelerator. The phantom is located the isocenter point, eccentric 3 cm, eccentric 6 cm, and different points on the z-axis successively on the accelerator. The scanned mode is the standard head mode. The reconstructive thickness is 2.5 cm. The different densities inserts of CTP404 module on the different locations are measured via Eclips treatment planning system (TPS) and computed by Matlab 7.0 and the CT linear fitting are then processed. In order to understand better the linear distribution along with the value of CT in the spatial distribution the results are compared with the fan-beam CT. Results Phantom studies show that:CBCT has good linearity performance not only under the standard header (body) of the scanning conditions, but also on such locations including the cross-sectional, the sagittal, the coronal plane and the eccentric position (R2>0.953). Bowtie filtration device does not change the CT linearity but changes the value of CT. Conclusions The linearity of X-ray CBCT on the Varian linear accelerator is favorable. CBCT will be used in the TPS dose calculation via further correction of the CT value.
ZHANG Jun*,XU Li-ming,LIU Hui et al. Evaluation of the linearity characteristic of the cone-beam CT fixed on the Varian 23EX linear accelerator[J]. Chinese Journal of Radiation Oncology, 2012, 21(5): 464-467.
ZHANG Jun*,XU Li-ming,LIU Hui et al. Evaluation of the linearity characteristic of the cone-beam CT fixed on the Varian 23EX linear accelerator[J]. Chinese Journal of Radiation Oncology, 2012, 21(5): 464-467.
[1] Lei X, Thorndkke B, Schreibmann E, et al. Overview of image guided radiation therapy. Int J Med Dosim,2006,31:91-112. [2] Yang Y, Li T, Schreibmann E, et al. Is cone beam CT suitable for dose verification? Annual meeting of American association of physicists in medicine, Seattle,USA,2005.New York:AAPM,2005. [3] Yang Y, Schreibmann E, Xing L, et al. Evaluation of kV cone-beam CT (CBCT)-based dose calculation. Phys Med Biol,2007,52:685-705. [4] 戴建荣,胡逸民.图像引导放射治疗的实现方式.中华放射肿瘤学杂志,2006,15:133-135. [5] Wang G, Lin TH, Cheng PC, et al. Error analysis on a generalized Feldkamp′s cone-beam computed tomography
algorithm. Int J Scanning,1995,17:361-370. [6] Grass M, Khler TH, Proksa R, et al. 3-D cone-beam CT reconstruction for circular trajectories. Int J Med Phys Biol,2000,45:329-347. [7] 邵义文,卢文婷.锥形束CT系统的散射校正方法分析.中国医学物理学杂志,2008,25:634-637. [8] Mail N, Moseley DJ, Siewerdsen JH, et al. The influence of bowtie filtration on cone-beam CT image quality. Med Phys,2009,36:22-32. [9] 姜伟,柏森,周亮强,等.Bowtie 滤过器对加速器锥形束CT图像质量影响.中华放射肿瘤学杂志,2010,19:548-551.