AbstractObjective To analyze the precision and stability of optical surface imaging system for patients who received radiotherapy with active breath control. Methods Eighteen radiotherapy patients with lung metastasis were managed by active breath control (ABC).The difference error detected by optical surface imaging system and CBCT were defined as the precision of optical surface imaging system. The variation among the error of optical surface imaging positioning the value of correction of treatment position and the error detected by optical surface imaging again were defined as the stability of optical surface imaging system. Intrafractional errors were analyzed by optical surface imaging system through whole treatment process (including breath hold and free breath). Results The optical surface imaging system had precision (systematic (Σ) and random errors (σ)) of 1.78/3.42 mm 2.54/6.57 mm and 2.79/3.22 mm respectively and stability of2.12/2.54 mm 3.09/4.02 mm and 1.37/3.55 mm respectively in lateral-medial superior-inferior and anterior-posterior directions. The intrafractional errors (Σ and σ) were 0.42/0.85 mm 0.41/1.47 mm and 0.41/1.47 mm respectively for breath hold duration and 4.76/4.16 mm 6.54/7.78 mm and 3.13/5.92 mm for free breath duration in lateral-medial superior-inferior and anterior-posterior directions. Conclusions As an effective method for validate breath hold;Optical surface imaging system can improve the precision and safety of active breath control.However,the factors that affect the accuracy and stability of the optical surface imaging system in patients undergoing radiotherapy with ABC are not clear; it cannot replace the CBCT for positioning verification.
Fund:Sichuan Science and Technology Department project (2016FZ0086)
Corresponding Authors:
Li GuangjunEmail ginick829@sina.com
Cite this article:
Zhong Renming,Ye Chengwei,Li Liqin et al. Analysis of utility of optical surface imaging system for patients who received radiotherapy with active breath control[J]. Chinese Journal of Radiation Oncology, 2018, 27(1): 89-93.
Zhong Renming,Ye Chengwei,Li Liqin et al. Analysis of utility of optical surface imaging system for patients who received radiotherapy with active breath control[J]. Chinese Journal of Radiation Oncology, 2018, 27(1): 89-93.
[1] Nabavizadeh N,Elliott DA,Chen YY,et al. Image guided radiation therapy (IGRT) practice patterns and IGRT′s impact on workflow and treatment planning:results from a national survey of American society for radiation oncology members[J].Int J Radiat Oncol Biol Phys,2016,94(4):850-857.DOI:10.1016/j.ijrobp.2015.09.035 [2] Cubillos Mesías M,Boda-Heggemann J,Thoelking J,et al. Quantification and assessment of interfraction setup errors based on cone beam CT and determination of safety margins for radiotherapy[J/OL].PLoS One,2016,11(3):e0150326.DOI:10.1371/journal.pone.0150326 [3] Carl J,Sander L.Five-year follow-up using a prostate stent as fiducial in image-guided radiotherapy of prostate cancer[J].Acta Oncol,2015,54(6):862-867.DOI:10.3109/0284186X.2014.987355 [4] Wang D,Zhang Q,Eisenberg BL,et al. Significant reduction of late toxicities in patients with extremity sarcoma treated with image-guided radiation therapy to a reduced target volume:results of radiation therapy oncology group RTOG-0630 Trial[J].J Clin Oncol,2015,33(20):2231-2238.DOI:10.1200/JCO.2014.58.5828 [5] Kaul D,Budach V,Graaf L,et al. Outcome of elderly patients with meningioma after image-guided stereotactic radiotherapy:a study of 100 cases[J].BioMed Res Int,2015,2015:868401.DOI:10.1155/2015/868401 [6] Hess CB,Thompson HM,Benedict SH,et al. Exposure risks among children undergoing radiation therapy:considerations in the era of image guided radiation therapy[J].Int J Radiat Oncol Biol Phys,2016,94(5):978-992.DOI:10.1016/j.ijrobp.2015.12.372 [7] Zhang YB,Wu H,Chen Z,et al. Concomitant imaging dose and cancer risk in image guided thoracic radiation therapy[J].Int J Radiat Oncol Biol Phys,2015,93(3):523-531.DOI:10.1016/j.ijrobp.2015.06.034 [8] Bujold A,Craig T,Jaffray D,et al. Image-guided radiotherapy:has it influenced patient outcomes?[J].Semin Radiat Oncol,2012,22(1):50-61.DOI:10.1016/j.semradonc.2011.09.001 [9] Zhong RM,Wang J,Jiang XQ,et al. Hypofraction radiotherapy of liver tumor using cone beam computed tomography guidance combined with active breath control by long breath-holding[J].Radiother Oncol,2012,104(3):379-385.DOI:10.1016/j.radonc.2011.11.007 [10] Stieler F,Wenz F,Shi M,et al. A novel surface imaging system for patient positioning and surveillance during radiotherapy. A phantom study and clinical evaluation[J].Strahlenther Onkol,2013,189(11):938-944.DOI:10.1007/s00066-013-0441-z [11] Zhong RM,Wang J,Zhou L,et al. Implementation of single-breath-hold cone beam CT guided hypofraction radiotherapy for lung cancer[J].Radiat Oncol,2014,9:77.DOI:10.1186/1748-717X-9-77 [12] Moser T,Habl G,Uhl M,et al. Clinical evaluation of a laser surface scanning system in 120 patients for improving daily setup accuracy in fractionated radiation therapy[J].Int J Radiat Oncol Biol Phys,2013,85(3):846-853.DOI:10.1016/j.ijrobp.2012.05.026 [13] Freislederer P,Reiner M,Hoischen W,et al. Characteristics of gated treatment using an optical surface imaging and gating system on an Elekta linac[J].Radiat Oncol,2015,10:68.DOI:10.1186/s13014-015-0376-x [14] Schönecker S,Walter F,Freislederer P,et al. Treatment planning and evaluation of gated radiotherapy in left-sided breast cancer patients using the CatalystTM/SentinelTM system for deep inspiration breath-hold (DIBH)[J].Radiat Oncol,2016,11:143.DOI:10.1186/s13014-016-0716-5 [15] Jönsson M,Ceberg S,Nordström F,et al. Technical evaluation of a laser-based optical surface scanning system for prospective and retrospective breathing adapted computed tomography[J].Acta Oncol,2015,54(2):261-265.DOI:10.3109/0284186X.2014.948059 [16] Wikström K,Nilsson K,Isacsson U,et al. A comparison of patient position displacements from body surface laser scanning and cone beam CT bone registrations for radiotherapy of pelvic targets[J].Acta Oncol,2014,53(2):268-277.DOI:10.3109/0284186X.2013.802836 [17] Pallotta S,Marrazzo L,Ceroti M,et al. A phantom evaluation of SentinelTM,a commercial laser/camera surface imaging system for patient setup verification in radiotherapy[J].Med Phys,2012,39(2):706-712.DOI:10.1118/1.3675973 [18] Moser T,Fleischhacker S,Schubert K,et al. Technical performance of a commercial laser surface scanning system for patient setup correction in radiotherapy[J].Phys Med,2011,27(4):224-232.DOI:10.1016/j.ejmp.2010.10.005 [19] Pallotta S,Vanzi E,Simontacchi G,et al. Surface imaging,portal imaging,and skin marker set-up vs. CBCT for radiotherapy of the thorax and pelvis[J].Strahlenther Onkol,2015,191(9):726-733.DOI:10.1007/s00066-015-0861-z [20] Gopan O,Wu QW.Evaluation of the accuracy of a 3D surface imaging system for patient setup in head and neck cancer radiotherapy[J].Int J Radiat Oncol Biol Phys,2012,84(2):547-552.DOI:10.1016/j.ijrobp.2011.12.004