The technique and development of total skin electron irradiation—Taiwan experiences
Wu Jiaming*,Ye Shian,Ren Yimin
*Department of Radiation Oncology,E-Da Hospital;Department of Medical Imaging and Radiological Sciences, I-Shou University, Medical Physics & Informatics Lab;Department of Electronic Engineering, National Kaohsiung University of Applied Science;Kaohsiung 84001,Taiwan,China
Abstract Objective The aim of this study is to present two novel technique-Lying on position and rotational board TSEI in which uniform dose to the whole body as well as the soles and vertex of scalp can be achieved in one electron beam treatment fraction. Methods Patients are treated with two groups in prone and supine position by leaning on inner rotational board in prone and supine position, each group can further be separated into two subgroup with tilting and rotation positions for treatment. By incorporating all these setting on the aswirl board with the optimal gantry angle, the patient will final accumulate compensated dosages on whole body surface. Results The dose uniformity for the rotational board-TSET or Lying on position could be control within ±10%. Conclusions The two TSEItechnique-Lying on and Rotational board TSEI techniques presented in this study is able to deliver a uniform dose to the patient’s skin surface as well as to the vertex and the soles all in one time, eliminating the troubles of having to further boost fields for these two regions when using the Stanford Six Field Technique.
Wu Jiaming*,Ye Shian,Ren Yimin. The technique and development of total skin electron irradiation—Taiwan experiences[J]. Chinese Journal of Radiation Oncology, 2015, 24(4): 449-453.
Wu Jiaming*,Ye Shian,Ren Yimin. The technique and development of total skin electron irradiation—Taiwan experiences[J]. Chinese Journal of Radiation Oncology, 2015, 24(4): 449-453.
[1] American association of physicists in medicine. Total skin electron therapy technique and dosimetry. Report No.23[S]. New York:American Institute of Physics,1988. [2] Peters VG.Use of an electron reflector to electron therapy[J]. Int J Radiat Oncol Biol Phys,2000,46(4):1065-1069. [3] Weaver RD, Gerbi BJ, Dusenbery KE. Evaluation of dose variation during total skin electron irradiation using thermoluminescent dosimeters[J]. Int J Radiat Oncol Biol Phys,1995,33(2):475-478. [4] Shouman T, Zeinabel T.Total skin electron therapy:a modified technique for small room linear accelerator[J]. J Egy Nat Can Inst,2004,16(4):202-209. [5] Wu JM, Leung SW.Lying-on position of total skin electron therapy[J]. Int J Radiat Oncol Biol Phys,1997,39(2):521-528. [6] Wu JM, Yeh SA, Hsiao KY,et al. A conceptual design of rotating board technique for delivering total skin[J].Med Phys,2010,37(4):1449-1458. [7] Ester CP, Francisco SD.Total skin electron therapy treatment verification:Monte Carlo simulation and beam characteristics of large non-standard electron fields[J]. Phys Med Biol,2003,48(17):2783-2796.