摘要 提出一种快速定位检修医用直线加速器和螺旋断层治疗系统“GAS”连锁故障的有效方法。采用带红外线的制冷剂泄漏检测仪结合肥皂水泡进行检漏;采用砂纸、锉刀轻轻打磨漏点及周围5 mm 宽的波导表面,再利用大功率150 W的电烙铁进行锡焊补漏。快速定位检测出泄漏点,补漏后的波导管使用12年未发现泄漏。有了这一方法,医用直线加速器和螺旋断层治疗系统“GAS”联锁故障的检修并不难。
Abstract:In this paper, an effective method was proposed for rapid positioning and repair of medical Clinac and helical tomotherapy “GAS” interlock fault. The infrared refrigerant leakage detector was combined with soap bubbles for leakuseage detection. The sandpaper and file were utilized to gently grind the leakage point and the surrounding 5 mm-wide waveguide surface,and then the high-power electric soldering iron of 150 W was adopted to make up for the leakage. The leakage point was detected by rapid positioning,and no leakage was found in the waveguide after 12 years of use. This approach can significantly reduce the difficulty in the repair of the Clinac and helical tomotherapy"GAS" interlock fault.
[1] 胡逸民.放射物理学[M].北京:原子能出版社,1999. Fu YM.Radiologic physics[M].Beijing:Atomic energy press,1999. [2] 杨绍洲,陈龙华,张树军.医用电子直线加速器[M].北京:人民军医出版社,2004. Yang SZ,Cheng LH,Zhang SJ.Medical electron linear accelerator[M].Beijing:People’ s military medical press,2004. [3] 顾广本.医用加速器[M].北京:科学出版社,2003. Gu GB.Medical accelerator[M].Beijing:Science press,2003. [4] 徐建铭.加速器原理[M].北京:科学出版社,1981. Xu JM.Accelerator principle[M].Beijing:Science press,1981. [5] 姚允国.电子直线加速器[M].北京:科学出版社,1986. Yao YG.Electron linear accelerator[M].Beijing:Science press,1986. [6] 徐慧军.现代肿瘤放射物理学[M].中国原子能出版社,2015. Xu HJ.The modern physics of radiation oncology[M].China atomic energy press,2015. [7] 王振洲,姚翔,李朝伟.医用电子直线加速器波导系统及维护要点[J].中华放射肿瘤学杂志,2011,20(1):22-22.DOI:10.3760/cma.j.issn.1004-4221.2011.01.008. Wang ZZ,Yao X,Li CW.Medical electron linear accelerator waveguide system and maintenance points[J].Chin J Radiat Oncol, 2011,20(1):22-22.DOI:10.3760/cma.j.issn.1004-4221.2011.01.008.