Department of Radiation Oncology,Zhongnan Hospital of Wuhan University,Wuhan 430071,China (Zhang J,Xie CH,Zhou YF,Bao ZR,Wang DJ,Chen C,Liu H);School of Physics and Technology,Wuhan University,Wuhan 430072,China (Huang XT)
Abstract: Objective KV-CBCT was utilized to evaluate the setup errors in lung SBRT with R624-SCF immobilization equipment,quantitatively analyze the percentage of all types of errors in the cumulative errors and unravel the main sources of setup errors. Methods The CBCT data weekly and QA data monthly from 32 patients diagnosed with lung neoplasms were collected to quantitatively analyze the setup errors. The cumulative errors were calculated by statistical model. The proportion and source of each type of setup error was analyzed. Results All 32 patients received a total of 420 times of CBCT.The setup errors of immobilization equipment in the lateral,supine-inferior,anterior-posterior directions were (0.03±0.72) mm,(0.73±1.16) mm and (-0.21±0.95) mm,respectively. The errors of tumor motion in three directions were (0.71±2.61) mm,(-0.80±2.60) mm and (0.075±1.77) mm,respectively. According to the calculation formula proposed by Vance Keeling,the proportion of the cumulative error was 54.55%,9.21% for immobilization equipment,12.97% for tumor motion,2.55% for couch sagging,5.70% for Gantry radiation isocenter,4.73% for Collimator radiation isocenter,4.61% for couch radiation isocenter and 5.70% for X-ray field isocenter,respectively. Conclusions The main factors of setup errors during SBRT treatment for lung cancer are setup random,tumor motion,immobilization equipment,couch sagging and machine isocenter. During radiotherapy,targeted control of tumor motion is of significance for minimizing the cumulative errors.
Zhang Jun,Huang Xiaotian,Xie Conghua et al. Quantitative analysis of setup errors in lung SBRT with R624-SCF immobilization equipment[J]. Chinese Journal of Radiation Oncology, 2018, 27(4): 396-400.
[1] 郭明芳,郭雷鸣,赵娅琴,等.锥形束CT研究肺癌放射治疗中的摆位误差[J].中国肿瘤临床,2009,36(15):848-852.DOI:10.3969/j.issn.1000-8179.2009.15.003. Guo MF,Guo LM,Zhao YQ,et al. Interfractional and Intrafractional Errors in Radiotherapy of Lung Cancer Using Cone-beam Computed Tomography (CBCT)[J].Chin J Clin Oncol,2009,36(15):848-852.DOI:10.3969/j.issn.1000-8179.2009.15.003. [2] 刘晖,张俊,钟亚华,等.自制体部固定装置在SBRT放疗中的应用分析[J].中华放射肿瘤学杂志,2013,22(6):501-503.DOI:10.3760/cma.j.issn.1004-4221.2013.06.021. Liu H,Zhang J,Zhong YH,et al. The analysis of clinical application of home-made immobilization device in SBRT[J].Chin J Radiat Oncol,2013,22(6):501-503.DOI:10.3760/cma.j.issn.1004-4221.2013.06.021. [3] Keeling V,Hossain S,Jin H,et al. Quantitative evaluation of patient setup uncertainty of stereotactic radiotherapy with the frameless 6D ExacTrac system using statistical modeling[J].J Appl Clin Med Phys.2016,17(3):111-127.DOI:10.1120/jacmp.v17i3.5959. [4] Gao,JF.Liu,XQ.Off-Isocenter winston-lutz test for stereotactic radiosurgery/stereotactic body radiotherapy[J].Int J Med Phys,Clin Engineer Radiat Oncol,2016,154-161.DOI.org/10.4236/ijmpcero.2016.52017. [5] Van HM.Errors and margins in radiotherapy[J].Semin Radiat Oncol,2004,14(1):52-64. [6] Mcgarry CK,Cosgrove VP,Fleming VA,et al. An analysis of geometric uncertainty calculations for prostate radiotherapy in clinical practice[J].Br J Radiol,2009,82(974):140. [7] Harrison AJL.Technical overview of geometric uncertainties in radiotherapy[J].Br Inst Radiol,2003(1):11-44.