中华放射肿瘤学杂志  2016, Vol. 25 Issue (3): 270-274    DOI: 10.3760/cma.j.issn.1004-4221.2016.03.016
物理·生物·技术 最新目录| 下期目录| 过刊浏览| 高级检索 [an error occurred while processing this directive]|[an error occurred while processing this directive]
三维超声图像引导系统质控评价体系建立
韦晓波, 胡强, 沐金明, 徐国平, 陈远, 程常海, 李齐林, 裴红蕾
213003 常州,苏州大学附属第三医院放射肿瘤科
Establishment of quality control evaluation system for 3D-ultrasound-based image-guided radiotherapy
Wei Xiaobo,Hu Qiang,Mu Jinming,Xu Guoping,ChenYuan,Cheng Changhai,Li qilin,Pei Honglei
Department of Radiation Oncology,Third Affiliated Hospital of Soochow University,Changzhou 213003,China
全文: PDF (1035 KB)   HTML (1 KB) 
输出: BibTeX | EndNote (RIS)      背景资料
摘要 目的 分析三维超声IGRT系统(3DUS-IGRT)日常工作的稳定性和精确度,为软组织肿瘤放疗的临床运用提供依据。方法 用专用校准模体对3DUS-IGRT系统进行连续 1年的校准与质控,探索超声引导设备日常质控的方法并监测其稳定性和精度。结果 模体位置误差在3DUS-IGRT系统的模拟定位工作站(Sim工作站)与图像引导工作站(Guide工作站)均≤1 mm。结论 三维超声IGRT系统在整套严格且正确的校准与质控支持下性能稳定,能为软组织肿瘤精确放疗提供新的影像引导方法。
服务
把本文推荐给朋友
加入我的书架
加入引用管理器
E-mail Alert
RSS
作者相关文章
韦晓波
胡强
沐金明
徐国平
陈远
程常海
李齐林
裴红蕾
关键词 三维超声图像引导放射疗法质量控制模体    
Abstract: Objective To analyze the stability and accuracy of the equipment for three-dimensional ultrasound-based image-guided radiation therapy (3DUS-IGRT) in daily practice, and to provide a basis for clinical application of radiotherapy for soft tissue tumors. Methods A specific calibration phantom was used for continuous calibration and quality control of the 3DUS-IGRT equipment in a year. The method for daily quality control of ultrasound-guided equipment was explored, and its stability and accuracy were monitored. Results The phantom position errors in both Sim and Guide stations of the 3DUS-IGRT equipment were within 1 mm. Conclusions The 3DUS-IGRT equipment has a stable performance with the support of a complete set of stringent and accurate calibration and quality control, which provides a new image-guided method for precise radiotherapy for soft tissue tumors.
Key wordsThree-dimensional ultrasound    Image-guided radiation therapy    Quality control    Phantom   
收稿日期: 2014-07-28     
通讯作者: 裴红蕾,Email:jtw316@sina.com   
引用本文:   
韦晓波,胡强,沐金明等. 三维超声图像引导系统质控评价体系建立[J]. 中华放射肿瘤学杂志, 2016, 25(3): 270-274.
Wei Xiaobo,Hu Qiang,Mu Jinming et al. Establishment of quality control evaluation system for 3D-ultrasound-based image-guided radiotherapy[J]. Chinese Journal of Radiation Oncology, 2016, 25(3): 270-274.
链接本文:  
http://journal12.magtechjournal.com/Jweb_fszlx/CN/10.3760/cma.j.issn.1004-4221.2016.03.016     或     http://journal12.magtechjournal.com/Jweb_fszlx/CN/Y2016/V25/I3/270
京公网安备11010502022167号
版权所有 © 2010《中华放射肿瘤学杂志》编辑部
地址:北京朝阳区潘学园南里17号 中国医学科学院肿瘤医院(100021)
电话:010-67700737,87788294 Email: cjron@cmaph.org
本系统由北京玛格泰克科技发展有限公司设计开发 技术支持:Support@magtech.com.cn