Abstract:Objective To study the physical characteristics of the OPEN stereotactic body radiotherapy system for the clinical application. Methods The 0.125cc ioniztion chamber, 160 mm polystyrene sphere model, Gafchromic EBT2 films and IBA film analysis software were used to evaluated the focus position tolerance, dose rate, repeatability, linear relation, penumbra and composite error of the OPEN stereotactic body gamma knife. We used the DTA method to verify the accuracy of dose distribution between the plans and measured value. Resualts The focus error was 0.36 mm, max dose rate tolerance 3%, linear relation error 2%, repeatability error 0.3%, composited error 2.5 mm. There was 90% pass rate when thedistance away from test point was less than 2 mm and the dose error was set less than 5%. Conclusions Parts of the test resualts were similar to the head gamma knife national protocal of OPEN stereotatic body gamma knife. The deliver dose distribution can meet the clinic need.
Cao Hongbin,Wu Xudong,Liu Guoqiang et al. Physical characteristics research of OPEN stereotactic body gamma knife[J]. Chinese Journal of Radiation Oncology, 2015, 24(6): 699-702.
[1]武汉康桥医学新技术有限公司,KLF-A型OPEN式立体定向放射治疗系统产品使用说明书[M] .武汉:武汉康桥医学新技术有限公司,2003. [2]周振山,王军良,段小萍.LUNATM-260型伽玛射线立体定向回转聚焦放疗机质量控制[J] .医疗卫生装备,2011,32(3):69-70.DOI:10.3969/j.issn.1003-8868.2011.03.028. [3]毛伊斌,陈勇.LGK剂量规划系统输入数据的测试与验证[J] .中华放射医学与防护杂志,2011,21(6):452-454.DOI:10.3760/cma.j.issn.0254-5098.2001.06.021. [4]Watanabe Y,Akimitsu T,Hirokawa Y,et al. Evaluation of dose delivery accuracy of Gamma Knife by polymer gel dosimetry[J] .J Appl Clin Med Phys,2005,6(3):133-142,DOI:10.1120/jacmp.2025.25343. [5]国家质量监督检验检疫总局.JJG 1013-2006,头部立体定向放射外科γ辐射治疗源检定规程[M] .北京:中国计量出版社,2007. [6]胡逸民,肿瘤放射物理学[M] .北京:原子能出版社,1999. [7]IAEA 277.Absorbed dose determination in photon and electron beams:an international code of practice[M] .Michigan:International Atomic Energy Agency,1997. [8]Heck B,Jess-Hempen A,Kreiner HJ,et al. Accuracy and stability of positioning in radiosurgery:long term results of the Gamma Knife system[J] .Med Phys,2007,34(4):1487-1495.DOI:10.1118/1.2710949. [9]Wilcox EE,Daskalov GM.Evaluation of GAFCHROMIC EBT film for CyberKnife dosimetry[J] .Med Phys,2007,34(6):1967-1974.DOI:10.1118/1.2734384. [10]Aldelaijan S,Devic S,Mohammed H,et al. Evaluation of EBT-2 model GAFCHROMICTM film performance in water[J] .Med Phys,2010,37(7):3687-3693.DOI:10.1118/1.3455713. [11]Andrés C,del Castillo A,Tortosa R,et al. A comprehensive study of the Gafchromic EBT2 radiochromic film,A comparison with EBT[J] .Med Phys,2010,37(12):6271-6278.DOI:10.1118/1.3512792. [12]TG Report 69.Radiographic film for megavoltage beam dosimetry[J] .Med Phys,2007,34(6):2239-2240. [13]Goetsch SJ,Murphy BD,Schmidt R,et al. Physics of rotating gamma systems for stereotactic radiosurgery[J] .Int J Radiat Oncol Biol Phys,1999,43(3):689-696.DOI:10.1016/S0360-3016(98)00428-3.