Abstract:Objective To evaluate the effect of the setup error on the dosimetric verification for the nasopharyngeal carcinoma patients and cervical carcinoma patients treated with volumetric modulated arc therapy (VMAT). Methds VMAT plans for 10 cervical cancer patients and 10 nasopharyngeal carcinoma patients were transplanted into the Delta4 phantom and calculate the dose,next,implement the treatment on the Varian iX linear accelerator. on the Varian iX linear accelerator. To simulate the setup error by moving the treatment couch in, out, up, down, left, right by 3 mm,5 mm,7 mm. Thereby study the effect of the setup error on the pass rate of the dose verification. Results The results for the dose distribution using the gamma evaluation method showed that the pass rate (3%/3 mm) was less than 90% when the setup error were greater than 3 mm and 5 mm for the nasopharyngeal carcinoma patients and the cervical carcinoma patients. The pass rate of head direction were (64.7±8.2)% and (63.3±3.6)% on setup error of 5 mm and 7 mm for nasopharyngeal carcinoma patients and cervical carcinoma patients, respectively. Conclusions Setup error has great effect on the dose verification of the VMAT plans, the greater of the setup error, the lower of the pass rate. The setup error of head direction is more sensitive than other directions especially.
Sun Xiaohuan,Tan Lina,Ma Kui et al. Effect of the setup error on the dosimetric verification for the volumetric modulated arc therapy[J]. Chinese Journal of Radiation Oncology, 2014, 23(6): 513-514.
[1] Ahmad S, Vlachaki MT, Teslow TN, et al. Impact of setup uncertainty in the dosimetry of prostate and surrounding tissues in prostate cancer patient treated with peacock/IMRT[J].Med Dosim,2005,30:1-7. [2] Siebers JV, Keall PJ, Wu Q, et al. Effect of patient setup errors on simultaneously integrated boost head and neck IMRT treatment plans[J].Int J Radiat Oncol Biol Phys,2005,63:422-33. [3] Hong TS, Tomé WA, Chappell RJ, et al. The impact of daily setup variations on head-and-neck intensity-modulated radiation therapy[J]. Int J Radiat Oncol Biol Phys,2005,61:779-788. [4] 石锦平,谢秋英,张利文,等.鼻咽癌调强放疗摆位误差对剂量分布影响的研究[J].实用癌症杂志,2010,25:394-396. [5] 王鑫,胡超苏,应红梅,等.摆位系统误差对鼻咽癌放疗剂量分布的影响[J].中国癌症杂志,2008,18:620-625. [6] 张利文,石锦平,谢秋英,等.等中心偏移误差对容积调强放疗计划的剂量影响[J].肿瘤预防与治疗,2012,25:160-163. [7] Ezzell GA, Galvin JM, Low D,et al. Guidance document on delivery, treatment planning, and clinical implementation of IMRT:report of the IMRT subcommittee of the AAPM radiation therapy committee[J].Med Phys,2003,30:2089-2115. [8] Dong L, Antolak J, Salehpour M, et al. Patient-specific point dose measurement for IMRT monitor unit verification[J].Int J Radiat Oncol Biol Phys,2003,56:867-877. [9] 李高峰,朱庙生,吴钦宏,等.逆向计划调强适形放射治疗的质量保证[J].中华放射肿瘤学杂志,2002,11:190-193. [10] Dische S, Saunders MI, Williams C, et al. Precision in reporting the dose given in a course of radiotherapy[J]. Radiother Oncol,1993,29:287-293. [11] 许峰,柏森,王瑾,等.用锥形束CT图像测量放疗摆位误差[J].中华放射肿瘤学杂志,2007,16:461-464.