Abstract:Objective To study the CT number and noise of HT MVCT on different dose rate and scanning thickness. Methds The CT number of different relative electron density were measured in the MVCT image of Cheesephantom with tissue substitute plugs scanned with different dose rate and slice thickness. The physical density corresponding to the CT number was plotted as the image value to density table (IVDT). The noise was measured in the MVCT image of Cheesephantom with solid water plugs scanned with different dose rate and slice thickness. Results There was a significant different of the CT number of the plugs with different dose rates (P=0.000),it shows a positive correlation between the varied CT number and density (R2=0.846), there is larger impact on the high density number. There was still a significant effect on the noise with different dose rate (P=0.000 density), the noise increase as the dose rate decrease. There was no significant effect on IVDT (P=1.000) and noise (P=0.667) with different slice thickness. Conclusions The CT number and the noise vary with the dose rate, the QA of MVCT should be performed regularly to assure the quality of image and the accuracy of dose calculating on MVCT in adaptive.
Yue Qi,Duan Jimei,Wang Zhiwei et al. The research on the factors of effecting the CT number and noise of TomoTherapy MVCT[J]. Chinese Journal of Radiation Oncology, 2014, 23(6): 527-529.
[1] Ruchala KJ, Olivera GH, Schloesser EA, et al. Megavoltage CT on a tomotherapy system[J].Phys Med Biol,1999,442597-442621. [2] Meeks SL, Harmon JF Jr, Langen KM, et al. Performance characterization of megavoltage computed tomography imaging on a helical tomotherapy unit[J]. Med Phys,2005,32:2673-2681. [3] Langen KM, Meeks SL, Poole DO, et al. The use of megavoltage CT (MVCT) images for dose recomputations[J]. Phys Med Biol,2005,50:4259-4276. [4] Shah AP, Langen KM, Ruchala KJ, et al.Patient dose from megavoltage computed tomography imaging[J].Int J Radiat Oncol Biol Phys,2008,70:1579-1587. [5] Ramsey CR, Langen KM, Kupelian PA, et al. A technique for adaptive image-guided helical tomotherapy for lung cancer[J]. Int J Radiat Oncol Biol Phys,2006,64:1237-1244. [6] 刘景鑫,杨海山.螺旋CT 图像噪声影响因素的实验分析[J].中华放射学杂志,2000,34:9-11. [7] 陈伟,李文政,唐友林.螺旋CT 图像噪声的评价[J].放射学实践,2003,18:535-536. [8] 冯国生,梁远,吴丹玲,等.CT值-相对电子密度转换曲线的影响因素分析[J].中华放射肿瘤学杂志,2012,21:281-284. [9] 祁振宇,黄劭敏,邓小武.放疗计划CT 值的校准检测及其影响因素分析[J].癌症,2006,25:110-114.