Abstract:Objective To investigate the accuracy and reliability of the image automatic matching using X-ray volumetric image (XVI) system with the accelerator Elekta Synergy and to compare the matches with different methods and area. Methods A simulation plan using the head and neck phantom was designed and sent to XVI system for CBCT scanning. During the scanning,the couch for certain distance was moved and the data of XVI image automatic Matching in different matching methods and area was collected. Results The result of XVI image automatic matching is consistent and correct to direction and distance which the couch had been moved (x:(0.11 ± 0.41) mm, y:(-0.04± 2.6) mm, z:(0.28 ± 0.74 mm)).There are not significant differences between the results that matching in different matching methods and different matching area. Conclusions The automatic image Matching of XVI is accuracy and high reliability in recognition of offset error. But there are some significant differences on the automatic image Matching in different matching methods and different matching area.
Xu Senkui,Yao Wenyan,Lin Chengguang. The study of error in the head and neck phantom using the X-ray volumetric image system of Elekta Synergy accelerator with CBCT[J]. Chinese Journal of Radiation Oncology, 2014, 23(5): 433-436.
[1] Uematsu M, Akira S, Kazuhiko T, et al. Focal,high dose,and fractionate d modified stereotactic radiotherapy for lung carcinoma patients[J].Cancer,1998,82:1062-1070. [2] Teh BS, Bastasch MD, Wheeler TM, et al. IMRT for prostate cancer:defining target volume based on correlated pathologic volume of disease[J].Int J Radiat Oncol Biol Phys,2003,56:184-191. [3] 张爱华,胡健.OBI系统CBCT引导下的腹部调强患者位置验证的应用分析[J].医械临床,2010,31:82-90. [4] Jaffray DA, Siewerdsen JH, Wong JW, et al. Flat-panel cone-beam computed tomography for image radiation therapy[J].Int J Radiat Oncol Biol Phys,2002,53:1337-1349. [5] 蒋琳,陈晓品,金丹,等.应用OBI系统分析胸部肿瘤调强放疗的摆位误差[J].激光杂志,2009,30:92-95. [6] Houweling AC, van der Meer S, van der Wal E, et al. Improved immobilization using an individual head support in head and neck cancer patients[J].Radiother Oncol,2010,96:100-103. [7] 马广栋,李承军,付敬国,等.不同解剖区域配准方法对鼻咽癌体位误差的影响[J].医械临床,2011,32:63-64. [8] Lee N, Puri DR, Blanco AI, et al. Intensity-modulated radiation therapy in head and neck cancers:an update[J].Head Neck,2007,9:387-400. [9] Eisbruch A. Intensity modulated radiotherapy of head and neck cancer:encouraging early results[J]. Int J Radiat Oncol Biol Phys,2002,53:1-3. [10] Benner S, Roentgen B. Television monitoring of a 30 MV X-ray beam[J]. Phys Med Biol,1962,7:29-33. [11] 沈海英,沈淡琦,李永春,等.电子射野影像装置用于监控放射治疗中初次摆位误差的研究[J]. 生物医学工程与临床,2010,14:303-305. [12] Kim GY, Pawlicki T, Le QT, et al. Linac-base on-board imaging feasibility and the dosimetric consequences of head roll in head and neck IMRT plans[J]. Med Dosim,2008,33:93-99. [13] Van Herk M. Errors and margins in radiotherapy[J]. Semin Radiat Oncol,2004,14:52-64.