Dosimetric effect of multi-leaf collimator transmission on intensity-modulated radiation therapy
YANG Rui-jie*, WANG Jun-jie, ZHANG Xi-le, SUN Hai-tao, LIU Xuan-feng, CHEN Bin
*Department of Radiation Oncology, Third Hospital of Peking University, Beijing 100191, ChinaCorresponding author:WANG Jun-jie, Email:junjiewang47@yahoo.com
Abstract:Objective To investigate the dosimetric effect of multi-leaf collimator (MLC) transmission on intensity-modulated radiation therapy (IMRT). Methods MLC transmission through the leaves and rounded ends were measured with ion chamber for Varian Trilogy linear accelerator with the X-ray of 6 MV and 10 MV. The intraleaf and interleaf transmission were also measured with the electronic portal imaging device of aS1000. 10 tumor patients treated with IMRT were used to evaluate the MLC parameters modeling in the Eclipse treatment planning system. Results The average transmissions of Millennium MLC were 1.6% and 1.8% for 6 MV and 10 MV X-rays. The transmission increased with the field size and depth of measurement. The transmissions at off axis position were a little lower than those on the central axis. The intraleaf transmissions measured with aS1000 were 0.8%—1.2% and 1.1%—1.6%, the interleaf transmission were 1.3%—1.9% and 1.6%—2.5% for 6 MV and 10 MV X-rays respectively. Modeling with the measured MLC parameters, the Eclipse treatment planning system could calculate the dose distribution accurately. The γ pass rate at 3%3 mm was above 95%, except for two patients with nasopharyngeal cancer with the pass rate of 93.6% and 94.5%. Conclusions The transmissions through the MLC leaves and the leaf ends contribute to the dose throughout the target significantly for IMRT. MLC transmission varied with the field size, depth of measurement and off axis position.
YANG Rui-jie*,WANG Jun-jie,ZHANG Xi-le et al. Dosimetric effect of multi-leaf collimator transmission on intensity-modulated radiation therapy[J]. Chinese Journal of Radiation Oncology, 2013, 22(2): 157-162.
[1] Zygmanski P, Rosca F, Kadam D, et al. Determination of depth and field size dependence of multileaf collimator transmission in intensity-modulated radiation therapy beams. J Appl Clin Med Phys,2007,8:2693. [2] Lorenz F, Nalichowski A, Rosca F, et al. Spatial dependence of MLC transmission in IMRT delivery. Phys Med Biol,2007,52:5985-5999. [3] 符贵山,戴建荣,徐英杰,等.调强放疗计划系统的剂量学特性测试方法.中华放射肿瘤学杂志,2007,16:455-460. [4] LoSasso T, Chui CS, Ling C. Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy. Med Phys,1998,25:1919-1927. [5] Varian medical system. Eclipse 10 inverse planning administration and physics, Rev. 6.1.1.Palo Alto:Varian medical system,2010:604-607. [6] Wasb E, Valen H. Dosimetric discrepancies caused by differing MLC parameters for dynamic IMRT. Phys Med Biol,2008,53:405-415. [7] Webster GJ, Rowbottom CG, Mackay RI. Development of an optimum photon beam model for head and-neck intensity-modulated radiotherapy. J Appl Clin Med Phys,2007,8:2711. [8] Losasso T. IMRT delivery performance with a varian multileaf collimator. Int J Radiat Oncol Biol Phys,2008,71 Suppl 1:85-88.