The study of two-dimensional dosimetric verifications of volumetric intensity-modulated arc therapy based on electronic portal imaging devices
ZHU Jin-han*, CHEN Li-xin, JIN Guang-hua, CAO Wu-fei, DENG Xiao-wu, LIU Xiao-wei
*Department of Radiation Oncology, Cancer Centre, Sun Yat-Sen University, Guangzhou 510060, China Corresponding author:CHEN Li-xin, Email:chenlx@sysucc.org.cn
Abstract:Objective To develop an accurate 2D dose reconstruction model using electronic portal imaging device (EPID). 2D dosimetric verifications of volumetric intensity-modulated arc therapy (VIMAT) were done using the model. And the results were compared to other dosimeters. Methods The EPID-based dose reconstruction model was using convolve, deconvolve and correction function. The dose profiles which were obtained by the ion chamber were used to determine the model parameters. A total of 12 VIMAT plans for the treatment of anatomical sites of various complexities were chosen. The results obtained from EPID were compared to other dosimeters and treatment planning system (TPS). The ion chamber was used to measure the central point absolute doses. Other dosimeters were used to measure the plane dose distributions. All dosimeters measured the dose at 10 cm depth. The results were analyzed using γ-evaluation method. Results Regarding absolute central point doses, the ion chamber results were within 1.5% of the EPID results. For the comparison to Seven29 and Matrixx, the average γ pass rates with 2% and 2 mm criteria were 98.9% and 99.8% respectively. For the dose distributions measured by EPID and calculated by TPS, the γ pass rates with 3% and 3 mm criteria were 99.9%. Conclusions The presented results which were obtained from the comparison of measured and calculated doses show the reliability of our EPID-based dose reconstruction model. With the model, EPID can be a reliable and fast tool for IMRT plan dosimetric verification. The model expanded to the 3D dosimetric verification in the uniform phantom will be considered as the next work.
ZHU Jin-han*,CHEN Li-xin,JIN Guang-hua et al. The study of two-dimensional dosimetric verifications of volumetric intensity-modulated arc therapy based on electronic portal imaging devices[J]. Chinese Journal of Radiation Oncology, 2012, 21(6): 550-553.
[1] McDermott LN, Louwe RJ, Sonke JJ, et al. Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device. Med Phys,2004,31:285-295. [2] Louwe RJ, McDermott LN, Sonke JJ, et al. The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes. Med Phys,2004,31:2989-2995. [3] Berry SL, Polvorosa CS, Wuu CS. A field size specific backscatter correction algorithm for accurate EPID dosimetry. Med Phys,2010,37:2425-2434. [4] Winkler P , Georg D. An intercomparison of 11 amorphous silicon EPIDs of the same type:implications for portal dosimetry. Phys Med Biol,2006,51:4189-4200. [5] McDermott LN, Nijsten SMJJG, Sonke JJ, et al. Comparison of ghosting effects for three commercial a-Si EPIDs. Med Phys,2006,33:2448-2451. [6] Rowshanfarzad P, Sabet M, O′Connor DJ, et al. Reduction of the effect of non-uniform backscatter from an E-type support arm of a Varian a-Si EPID used for dosimetry. Phys Med Biol,2010,55:6617-6632. [7] Siebers JV, Kim JO, Ko L, et al. Monte Carlo computation of dosimetric amorphous silicon electronic portal images. Med Phys,2004,31:2135-2146. [8] Parent L, Seco J, Evans PM, et al. Monte Carlo modelling of a-Si EPID response:the effect of spectral variations with field size and position. Med Phys, 2006, 33:4527-4540. [9] Parent L, Fielding AL, Dance DR, et al. Amorphous silicon EPID calibration for dosimetric applications:comparison of a method based on Monte Carlo prediction of response with existing techniques. Phys Med Biol,2007,52:3351-3368. [10] Wang S, Gardner JK, Gordon JJ, et al. Monte Carlo-based adaptive EPID dose kernel accounting for different field size responses of imagers. Med Phys,2009,36:3582-3595. [11] Warkentin B, Steciw S, Rathee S, et al. Dosimetric IMRT verification with a flat-panel EPID. Med Phys,2003,30:3143-3155. [12] Jorgensen MK, Hoffmann L, Petersen JB, et al. Tolerance levels of EPID-based quality control for volumetric modulated arc therapy. Med Phys,2011,38:1425-1434. [13] Bakhtiari M, Kumaraswamy L, Bailey DW, et al. Using an EPID for patient-specific VMAT quality assurance. Med Phys,2011,38:1366-1373. [14] Wendling M, Louwe RJ, McDermott LN, et al. Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method. Med Phys,2006,33:259-273.