Abstract:Objective To evaluate setup errors of image guided radiation therapy (IGRT) for body tumors immbilized with vacuum cushions and localized with final isocenter marked method using kilovoltage cone beam CT (KVCBCT). Methods A retrospective study has been carried out for 223 patients from March 2009 to April 2011. All patients were immobilized with vacuum cushions and localized with final isocenter marked method in Philips PQS CT or Philips BrillianceCT Big Bore scanner, which were equipped with LAP movable laser systems. The CT images were transferred to a Varian Eclipse workstation for contouring and treatment planning. Before irradiation, a KVCBCT scan was performed and image registration was done on a Varian iX linear accelerator via OBI system. Each set of setup errors in right-left (RL), superior-inferior (SI), and anterior-posterior (AP) directions were gathered. An independent-samples t-test statistical analysis was conducted with the 758 sets of data using SPSS 16.0. Results The statistical analysis showed that setup errors from the 758 datasets were depicted a Gaussian distribution. The system errors ± random errors in RL, SI and AP were (-0.5±2.8) mm,(0.0±3.0) mm and (0.4±3.4) mm, respectively. Referring to the formula for planning target volume margin calculation, M=2.5Σ+0.7δ, the margins were calculated as 3.2, 2.1 and 3.4 mm, respectively. Conclusions The margins derived from this retrospective study have confirmed the premise that the treatment plans were executed in patients with high reliability, thereby created a high sense of confidence for the clinicians.
WU Qin-hong,LI Gao-feng,WANG Chan et al. Analysis of setup errors on CBCT-based image guided radiation therapy for body tumors patients[J]. Chinese Journal of Radiation Oncology, 2012, 21(1): 68-71.
[1] Mutic S, Palta JR, Butker EK,et al. Quality assurance for computed-tomography simulators and the computed-tomography-simulation process:report of the AAPM Radiation Therapy Committee Task Group No. 66.Med Phys,2003,30:2762-2792. [2] van Herk M, Remeijer P, Rasch C, et al. The probability of correct target dosage:dose-population histograms for deriving treatment margins in radiotherapy. Int J Radiat Oncol Biol Phys,2000,47:1121-1135. [3] van Herk M. Errors and margins in radiotherapy. Semin Radiat Oncol,2004,14:52-64. [4] International commission on radiation units and measurements. ICRU Report 62.Prescribing, recording, and reporting photon beam therapy (Supplement to ICRU Report 50).Bethesda:ICRU,1999. [5] Stroom JC, de Boer HC, Huizenga H, et al. Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability. Int J Radiat Oncol Biol Phys,1999,43:905-919. [6] Stroom JC, Heijmen BJ. Geometrical uncertainties, radiotherapy planning margins, and the ICRU-62 report. Radiother Oncol,2002,64:75-83. [7] Herman MG. Clinical use of electronic portal imaging. Semin Radiat Oncol,2005,15:157-167. [8] 戴建荣,胡逸民.图像引导放疗的实现方式.中华放射肿瘤学杂志,2006,15:132-135. [9] 刘跃平,刘新帆,李晔雄,等. 盆腔肿瘤三维适形放疗摆位重复性研究. 中华放射肿瘤学杂志,2006,15:313-316. [10] 邬筱莹,章真,王文超,等.电子射野影像装置对盆腔肿瘤放疗摆位误差的测定. 中华放射肿瘤学杂志,2007,16:52-54. [11] 张连胜,张寅,李明辉,等.用锥形束CT技术测量热塑成型膜固定患者的放疗摆位误差. 中华放射肿瘤学杂志,2008,17:219-222. [12] 洪金省,张纬建,陈金梅,等.胸段食管癌三维适形放疗摆位误差研究. 中华放射肿瘤学杂志,2009,18:182-185. [13] 吴君心,谢志原,林育毅,等.应用锥形束CT对盆腔肿瘤放疗计划靶区外放距离的研究. 中华放射肿瘤学杂志,2010,19:328-330. [14] 张寅,张连胜,肖建平,等.用图像引导技术提高胸腹部肿瘤大分割放疗的治疗精度. 中华放射肿瘤学杂志,2008,17:441-444. [15] 李炯雁,郭小毛,姚伟强,等.在线千伏级锥形束CT引导前列腺癌调强放疗摆位误差研究. 中华放射肿瘤学杂志,2010,19:541-543. [16] Guan H, Hammoud R, Yin FF. A positioning QA procedure for 2D/2D (kV/MV) and 3D/3D (CT/CBCT) image matching for radiotherapy patient setup. J Appl Clin Med Phys,2009,10:273-280. [17] Du W, Yang JN, Chang EL, et al. A quality assurance procedure to evaluate cone-beam CT image center congruence with the radiation isocenter of a linear accelerator. J Appl Clin Med Phys,2010,11:15-26.