Abstract:In this article, a case for random signal interference in resolver (or encoder) of carousel on Varian VitalBeam medical linear accelerator was illustrated. This case was considered as a rare failure of the technology platform (including TrueBeam) by Chinese, European and American technical experts in Varian. In this article, the differences of the carousel in the structure and control system between VitalBeam and traditional accelerator were comparatively analyzed to enable the peers to have a deeper understanding of resolver. The logic and methods to solve the fault were summarized, providing reference experience for analyzing the problems related to carousel. It is found that the interlocking system is imperfect, which provides a factual and technological basis for Varian to optimize the system.
Zhang Qingzhao,Ye Yufeng,Chen Hanwei et al. The case and thinking for signal interference in resolver of carousel on Varian VitalBeam medical linear accelerator[J]. Chinese Journal of Radiation Oncology, 2022, 31(11): 1052-1054.
[1] 杨绍洲,陈龙华,张树军.医用电子直线加速器[M].北京:人民军医出版社,2004:162. Yang SZ, Chen LH, Zhang SJ.Medical Linear Accelerator[M].Beijing: People's Military Medical Press,2004:162. [2] 国家癌症中心/国家肿瘤质控中心.医用电子直线加速器质量控制指南[J].中华放射肿瘤学杂志,2020,29(4):241-258. DOI: 10.3760/cma.j.cn113030-20200212-00001. National Cancer Center / National Cancer Quality Control Center. Quality control guidelines for medical electronic linear accelerators[J]. Chin J Radiat Oncol,2020,29(4):241-258. DOI: 10.3760/cma.j.cn113030-20200212-00001. [3] 邵健. 瓦里安Trilogy加速器Carousel工作原理及故障分析[J].中国医疗设备,2019,34(6):177-179. DOI: 10.3969/j.issn.1674-1633.2019.06.049. Shao J.Analysis of the fault and operational principle of carousel on Varian trilogy linear accelerator[J]. Chin Med Dev,2019,34(6):177-179. DOI: 10.3969/j.issn.1674-1633.2019. 06.049. [4] 杨绍洲, 王胜军. 加速器过滤器旋转托盘的控制原理和检修方法[J].医疗卫生装备,2005,26(9):70-70. DOI: 10.3969/j.issn.1003-8868.2005.09.044. Yang SZ, Wang SJ.Control principle and trouble shooting of filter carouse for medical electronic linear accelerators[J]. Chin Med Equip J, 2005,26(9):70-70. DOI: 10.3969/j.issn.1003-8868.2005.09.044. [5] 杨绍洲,沈庆贤. 2100C加速器CARR和FOIL联锁的检修方法[J].北京生物医学工程,2004,23(2): F003. DOI: 10.3969/j.issn.1002-3208.2004.02.024. Yang SZ, Shen QX.Maintenance method of CARR and FOIL interlock for 2100C medical electronic linear[J]. Beijing Biomed Eng, 2004,23(2): F003. DOI: 10.3969/j.issn.1002-3208.2004.02.024. [6] 李明,安书董,段宇博. 一种基于AD2S1210的旋转变压器位置解码及监控方法[J].信息技术与信息化, 2021(4):139-140. DOI: 10.3969/j.issn.1672-9528.2021.04.041. Li M, An SD, Duan YB.A position decoding and monitoring method based on resolver AD2S1210[J]. Information Technology & Informatization, 2021(4):139-140. DOI: 10.3969/j.issn.1672-9528.2021.04.041. [7] 高枫. Varian 2300C/D医用直线加速器HWFA连锁故障的检修[J].中国医疗器械杂志,2005,29(2):155-156. DOI: 10.3969/j.issn.1671-7104.2005.02.023. Gao F.A trouble shooting of HWFA interlock for Varian 2300C/D medical electronic linear accelerator[J]. Chin J Med Instrum, 2005,29(2):155-156. DOI: 10.3969/j.issn.1671-7104.2005. 02.023. [8] 宋伟平,周振远,林鹏. Varian 23cx直线加速器CARR和FOIL故障维修一例[J].医疗装备,2015,28(1):99-100. DOI: 10.3969/j.issn.1002-2376.2015.01.060. Song WP, Zhou ZY, Lin P.Troubleshooting of FOIL&CARR interlock for Varian 23cx medical linear accelerator[J]. Med Equip, 2015,28(1):99-100. DOI: 10.3969/j.issn.1002-2376. 2015.01.060. [9] 杨智祥, 季智勇, 杨爱建. Varian CL23EX直线加速器CARR连锁故障分析与检修[J].医疗卫生装备,2013,34(10):139-140. DOI:10.7687/J.ISSN.1003-8868.2013.10.139. Yang ZX, Ji ZY, Yang AJ.Analysis and maintenance of CARR interlock for Varian CL23EX medical electronic linear accelerator[J]. Chin Med Equip J, 2013, 34(10):139-140. DOI:10.7687/J.ISSN.1003-8868.2013.10.139. [10] 季智勇, 杨爱建, 凌金城. 瓦里安CL23EX加速器电缆磨损引发故障分析与检修[J].中国医疗设备,2013,28(6):141-142. DOI: 10.3969/j.issn.1674-1633.2013.06.055. Ji ZY, Yang AJ, Lin JC.Troubleshooting and analysis of Varian CL23EX accelerator[J]. Chin Med Dev,2013,28(6):141-142. DOI: 10.3969/j.issn.1674-1633.2013.06.055. [11] 薛开昶, 陈强, 罗宗鑫, 等. 磁阻旋转变压器位置检测的电磁干扰抑制方法[J].微电机,2021,54(3):73-77. DOI: 10.3969/j.issn.1001-6848.2021.03.014. Xue KC, Chen Q, Luo ZX, et al.EMI suppression method of position detection for variable reluctance resolver[J]. Micromotors, 2021,54(3):73-77. DOI: 10.3969/j.issn.1001- 6848.2021.03.014.