Abstract:Objective To evaluate the dosimetric effects of set-up errors on nasal NK/T cell lymphoma by introducing set-up errors into the radiotherapy planning system for dose reconstruction. Methods Ten patients with nasal NK/T cell lymphoma were recruited. A non-coplanar volumetric modulated arc therapy plan was designed for CT image and clinical target area of each patient. After the completion of the plan, the set-up errors were introduced into the radiotherapy plan by changing the ISO of the treatment, and dose calculation was performed to reconstruct the dose distribution. Results With the increase of system set-up errors, the dose of target was decreased and the order affected by set-up errors in different directions was:left-right direction> head-foot direction> front-rear direction. When the translational set-up errors in each direction were -3mm to 3 mm and the rotating set-up errors were -3° to 3°, the range of dose change in all targets was less than ±3%. When the set-up errors in all directions were ≤ 3mm,the dose of organ at risk was less than or similar to the prescribed dose. When the set-up errors were> 3mm, the doses of lens, spinal cord, parotid gland and optic nerve gradually exceeded the prescribed dose. Only when the rotating set-up errors were ≥ 3°, the dose of lens exceeded the prescribed dose. Special attention should be paid to the influence of the greater set-up errors in the left and right direction on lens, spinal cord and parotid gland, as well as on the spinal cord due to the larger set-up errors in the front and rear direction. After the actual set-up errors were introduced from our department, it exerted slight effect on the irradiation dose of GTV and CTV, which was less than ±2%. In a few cases, the dose of organ at risk potentially exceeded the prescribed dose limit, and special attention should be diverted to overdose of the lens and optic nerve. Conclusions The set-up errors will result in target dose deficiency and overdose of organ at risk in nasal NK/T cell lymphoma, especially upon the set-up errors in the left and right direction. The effect of 3mm and 3° set-up errors on target and organ at risk is limited. Therefore, it is recommended to maintain the single direction set-up errors within 3mm and 3°. The actual set-up errors introduced from our department exert little effect on the target dose, but a small number of organs are at risk of exceeding the prescribed dose limit. It is necessary to increase the evaluation of the extension region of organ at risk.
Wang Zhanyu,Tan Junwen,Long Yusong et al. Dosimetric effect of set-up errors on nasal NK/T cell lymphoma based on dose reconstruction[J]. Chinese Journal of Radiation Oncology, 2019, 28(11): 849-853.
[1] van Herk M. Errors and margins in radiotherapy[J]. Semin Radiat Oncol,2004,14(1):52-64. DOI:10.1053/j.semradonc.2003.10.003. [2] 徐鹭英,潘建基,王笑良,等. 头颈部肿瘤图像引导调强放疗中摆位误差对剂量学的影响研究[J]. 中华放射肿瘤学杂志,2011,20(6):506-509. DOI:10.3760/cma.j.issn.1004-4221.2011.06.017. Xu LY,Pan JJ,Wang XL,et al. The preliminary study of setup errors′ impact on dose distribution of image guide radiation therapy for head and neck cancer[J]. Chin J Radiat Oncol,2011,20(6):506-509. DOI:10.3760/cma.j.issn.1004-4221.2011.06.017. [3] 张彪,贾鹏飞,汤乐民. 摆位稳健性对IMRT与VMAT剂量学影响研究[J]. 中华放射肿瘤学杂志,2016,25(7):676-680. DOI:10.3760/cma.j.issn.1004-4221.2016.07.003. Zhang B,Jia PF,Tang LM. Effects of positioning robustness on dosimetry for intensity-modulated radiation therapy and volumetric modulated arc therapy[J]. Chin J Radiat Oncol,2011,20(6):506-509. DOI:10.3760/cma.j.issn.1004-4221.2016.07.003. [4] 沈君姝,朱锡旭,李兵,等. 鼻咽癌调强放疗中摆位误差对物理剂量学的影响[J]. 中国医学物理学杂志,2008,25(2):552-554. DOI:1005-202X (2008)02-0552-03. Shen JS,Zhu XX,Li B,et al. Effects of the setup error on the physics dose for nasopharyngeal carcinoma undergoing the intensity modulated radiation therapy[J]. Chin J Med Phys,2008,25(2):552-554. DOI:1005-202X (2008)02-0552-03. [5] 张一戈,邱小平,陈维军,等. 基于图像引导技术对头颈部肿瘤临床摆位误差的剂量学影响[J]. 中国医学物理学杂志,2015,32(1):25-30. DOI:1005-202X (2015)01-25-06. Zhang YG,Qiu XP,Chen WJ,et al. Dosimetry influence study on head & Neck cancer clinical placement error based on image-guided radiotherapy technique[J]. Chin J Med Phys,2015,32(1):25-30. DOI:1005-202X (2015)01-25-06.