Commissioning of dynamic jaw delivery in tomotherapy
Hu Zhihui, Zhang Ke, Huang Peng, Dai Jianrong
Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital, Chinese Academy of Medical Sciences ( CAMS ) and Peking Union Medical College (PUMC), Beijing 100021,China
Abstract:Objective To evaluate the dosimetric penumbra and delivery accuracy of dynamic jaw delivery in tomotherapy. Methods The jaw positioning hardware and the beam model were updated. Mechanical alignments were verified after the upgrade of the jaw positioning hardware. PDDs and beam profiles were measured by a two-dimensional water tank and compared with the new beam model. Dose penumbras in the longitudinal direction were compared between the dynamic and fixed jaw plans for different field width. Delivery accuracy was evaluated by point dose measurements with A1SL chamber and gamma analysis on the dose distribution measured by ArcCheck detector array. Results Mechanical alignments were in tolerance and beam characteristics were tuned to match the dynamic jaw beam model. Differences in the field width between the measured results and reference data were< 0.3% for both symmetric and asymmetric profiles in the longitudinal direction. The dose penumbra in dynamic jaw delivery was reduced from 17.92 mm to 7.51 mm for 2.5 cm jaw, and from 33.73 mm to 6.97 mm for 5.0 cm jaw, close to the penumbra of the traditional 1.0 cm jaw. IMRT verification of clinical cases was performed by A1SL ion chamber and ArcCheck detector array. The mean point dose difference was 0.33%±0.73% between the calculated and meassured data. Gamma analysis of dose distributions revealed that approximately 99.8% of the points satisfied the gamma criteria of 3% dose difference and 3 mm distance-to-agreement and the mean passing rates remained 97.9% even with tightest criteria of 2%/2 mm, and 100.0% with the criteria of 4%/4 mm, respectively. Conclusions Dosimetric penumbra in the longitudinal direction is significantly improved by the dynamic jaw delivery. Both the mechanical alignment and treatment delivery are qualified, suggesting that this new treatment is accurate and reliable.
Hu Zhihui,Zhang Ke,Huang Peng et al. Commissioning of dynamic jaw delivery in tomotherapy[J]. Chinese Journal of Radiation Oncology, 2019, 28(2): 125-130.
[1] Chen Y,Chen Q,Chen M,et al. Dynamic tomotherapy delivery[J]. Med Phys,2011,38(6):3013-3024.DOI:10.1118/1.3584198. [2] Rong Y,Chen Y,Shang L,et al. Helical tomotherapy with dynamic running-start-stop delivery compared to conventional tomotherapy delivery[J]. Med Phys,2014,41(5):051709.DOI:10.1118/1.4870987. [3] Sterzing F,Uhl M,Hauswald H,et al. Dynamic jaws and dynamic couch in helical tomotherapy[J]. Int J Radiat Oncol Biol Phys,2010,76(4):1266-73.DOI:10.1016/j.ijrobp.2009.07.1686. [4] Manabe Y,Shibamoto Y,Sugie C,et al. Helical and static-port tomotherapy using the newly-developed dynamic jaws technology for lung cancer[J]. Technol Cancer Res Treat,2015,14(5):583-91.DOI:10.7785/tcrtexpress.2013.600280. [5] Sugie C,Manabe Y,Hayashi A,et al. Efficacy of the Dynamic Jaw Mode in Helical Tomotherapy With Static Ports for Breast Cancer[J]. Technol Cancer Res Treat,2015,14(4):459-65.DOI:10.1177/1533034614558746. [6] Langen KM,Papanikolaou N,Balog J,et al. QA for helical tomotherapy:Report of the AAPM Task Group 148[J]. Med Phys,2010,37(9):4817–53.DOI:10.1118/1.3462971. [7] Lee FK,Chan SK,Chau RM. Dosimetric verification and quality assurance of running-start-stop (RSS) delivery in tomotherapy[J]. J Appl Clin Med Phys,2015,16(6):23-29.DOI:10.1120/jacmp.v16i6.5336. [8] Low DA,Harms WB,Mutic S,et al. A technique for the quantitative evaluation of dose distributions[J]. Med Phys,1998,25(5):656–61.DOI:10.1118/1.598248. [9] Katayama S,Haefner MF,Mohr A,et al. Accelerated tomotherapy delivery with TomoEdge technique[J]. J Appl Clin Med Phys,2015,16(2):4964.DOI:10.1120/jacmp.v16i2.4964. [10] Krause S,Beck S,Schubert K,et al. Accelerated large volume irradiation with dynamic Jaw/Dynamic Couch Helical Tomotherapy[J]. Radiat Oncol,2012,7:191.DOI:10.1186/1748-717X-7-191. [11] Fetzner L,Aynsley E,Beck S,et al. Lung and liver SBRT using helical tomotherapy-a dosimetric comparison of fixed jaw and dynamic jaw delivery[J]. J Appl Clin Med Phys,2014,15(3):4664–71.DOI:10.1120/jacmp.v15i3.4664. [12] Lu W. A non-voxel-based broad-beam (NVBB) framework for IMRT treatment planning[J].Phys Med Biol,2010,5(23):7175-210.DOI:10.1088/0031-9155/55/23/002. [13] Lu W,Chen M. Fluence-convolution broad-beam (FCBB) dose calculation[J]. Phys Med Biol,2010,55(23):7211-29.DOI:10.1088/0031-9155/55/23/003. [14] Chen Q,Lu W,Chen Y,et al. Validation of GPU based TomoTherapy dose calculation engine[J]. Med Phys,2012,39(4):1877-86.DOI:10.1118/1.3693057.