Abstract:Objective To evaluate the clinical application of Catalyst system in positioning patients during cervical cancer radiotherapy, and to analyze its correlation with the bladder volume and body mass index (BMI) of patients. Methods A total of 33 patients diagnosed with cervical cancer from July to December 2017 in our hospital were included in the study. All patients were auxiliary positioned by an optical surface imaging system (C-Rad Catalyst) before each treatment. The CBCT imaging was executed twice a week. The setup errors from Catalyst and CBCT in the anterior-posterior (AP), superior-inferior (SI) and left-right (LR) directions were recorded. Each patient′s bladder volume and BMI were also recorded. Results The setup errors between Catalyst with masks and CBCT had the significant difference in SI (P<0.05) and LR (P<0.05). For Catalyst without masks, the setup errors with the bladder volume of 200–300 ml had the significant association in SI (R=-0.316, P<0.05). For the bladder volume of>300 ml, the setup errors for Catalyst with masks had the significant association in AP (R=-0.493, P<0.05), and that without masks had the significant association in SI and LR (R=0.335, P<0.05, R=-0.348, P<0.05). For patients of<25 kg/m2, setup errors for Catalyst with masks had the significant association with the BMI in LR (R=0.197, P<0.05);for ≥25 kg/m2, that with masks had the significant association in AP and SI (R=0.818, P<0.05;R=-0.498, P<0.05), that without masks had the significant association in AP and LR (R=0.652, P<0.05;R=-0.558, P<0.05). Conclusion Unlike CBCT system, the patient positioning by Catalyst system was easily affected by the bladder volume and BMI of patients.
[1] Quigley MM,Mate TP,Sylvester JE.Prostate tumor alignment and continuous,real-time adaptive radiation therapy using electromagnetic fiducials:clinical and cost-utility analyses[J].UrolOncol,2009,27(5):473-482.DOI:10.1016/j.urolonc.2008.04.017. [2] 屈超,梁广立.热塑体膜固定装置在宫颈癌调强放疗中的临床应用价值[J].医疗卫生装备,2015,36(12):92-93.DOI:10.7687/J.ISSN1003-8868.2015.12.092. Qu C,Liang GL.Clinical application value of thermoplastic body membrane fixation device in intensity modulated radiotherapy of cervical cancer[J].Chin Med Equip J,2015,36(12):92-93.DOI:10.7687/J.ISSN1003-8868.2015.12.092. [3] Wikstrm K,Nilsson K,Isacsson U,et al. A comparison of patient position displacements from body surface laser scanning and cone beam CT bone registrations for radiotherapy of pelvic targets[J].Acta Oncologica,2014,53(2):268-277.DOI:10.3109/0284186X.2013.802836. [4] Torsten M,Gregor H,Matthias U,et al. Clinical evaluation of a laser surface scanning system in 120 patients for improving daily setup accuracy in fractionated radiation therapy[J].Int J Radiation Oncol Biol Phys,2013,85(3):846-853.DOI:10.1016/j.ijrobp.2012.05.026. [5] Philipp F,Michael R,et al. Characteristics of gated treatment using an optical surface imaging and gating system on an Elekta linac[J].Radiat Oncol,2015,68(10):1748-717X.DOI:10.1186/s13014-015-0376-x. [6] Pallotta S,Russo S,Esposito M,et al. Testing the C-RAD Catalyst elastic image registration software using a deformable female phantom[J].Radiother Oncol,2015,115:520-521.DOI:10.1016/S0167-8140(15)40970-3. [7] 罗焕丽,彭海燕,靳富,等.Catalyst系统在乳腺癌放疗摆位中的应用及影响因分析[J].中华放射肿瘤学杂志,2018,27(2):190-194.DOI:10.3760/cma.j.issn.1004-4221.2018.02.014. Luo HL,Peng HY,Jing F,et al. Application and influence of catalyst system in radiotherapy of breast cancer[J].Chin J Radiat Oncol,2018,27(2):190-194.DOI:10.3760/cma.j.issn.1004-4221.2018.02.014. [8] Manfred W,Kathrin K,Lutz L.Clinical validation of two surface imaging systems for patient positioning in percutaneous radiotherapy[M].NY:Physics,2016. [9] Scherrer D,Bernhardt M,Lohr F.et al Clinical evaluation of a commercial surface-imaging system for patient positioning in radiotherapy[J].Strahlenther Onkol,2012,188(12):1080-1084.DOI:10.1007/s00066-012-0244-7. [10] Stieler F,Wenz F,Shi M.et al. A novel surface imaging system for patient positioning and surveillance during radiotherapy[J].Strahlenther Onkol,2013,189(11):938-944.DOI:10.1007/s00066-013-0441-z. [11] Pallotta S,Vanzi E,Simontacchi G,et al. Surface imaging,portal imaging,and skin marker set-up vs. CBCT for radiotherapy of the thorax and pelvis[J].Strahlenther Onkol,2015,191(9):726-733.DOI:10.1007/s00066-015-0861-z. [12] 罗焕丽,靳富,王颖,等.肿瘤放疗中体外误差与体内误差的关联性[J].中国医学物理学杂志,2016,33(1):10-15.DOI:10.3969/j.issn.1005-202X.2016.01.003. Luo HL,Jin F,Wang Y,et al. Correlation between in vitro error and in vivo error in tumor radiotherapy[J].Chin J Med Phys,2016,33(1):10-15.DOI:10.3969/j.issn.1005-202X.2016.01.003. [13] Walter F,Freislederer P,Belka C,et al. Evaluation of daily patient positioning for radiotherapy with a commercial 3D surface-imaging system (CatalystTM)[J].Radiat Oncol,2016,154(11):1748-717X.DOI:10.1186/s13014-016-0728-1. [14] Wikstrm K,Nilsson K,Isacsson,et al. A comparison of patient position displacements from body surface laser scanning and cone beam CT bone registrations for radiotherapy of pelvic targets[J].Acta Oncol,2014,53(2):268-277.DOI:10.3109/0284186X.2013.802836. [15] 何英剑,陶秋山,李晓婷,等.性别、年龄对成人超重与肥胖诊断标准影响[J].中国公共卫生,2009,25(4):441-443.DOI:10.3321/j.issn:1001-0580.2009.04.021. He YJ,Tao QS,Li XT,et al. Effects of gender and age on diagnostic criteria for overweight and obesity in adults[J].Pub Heal Chin,2009,25(4):441-443.DOI:10.3321/j.issn:1001-0580.2009.04.021. [16] 罗焕丽,王颖,李芳,等.便捷式膀胱容量测定仪 BVI9400及超声系统 iU22 评估膀胱容量的对比研究[J].中国医疗器械杂志,2015,39(4):298.DOI:10.3969/j.issn.1671-7104.2015.04.017. Luo HL,Wang Y,Li F,et al. Comparative study on the evaluation of bladder capacity by BVI9400 and ultrasonic system iU22 of convenient bladder capacity tester[J].Chin J Med Dev,2015,39(4):298.DOI:10.3969/j.issn.1671-7104.2015.04.017. [17] Brahme A,Nyman P,Skatt B.4D laser camera for accurate patient positioning,collision avoidance,image fusion and adaptive approaches during diagnostic and therapeutic procedures[J].MedPhys,2008,35(16):70-81.DOI:10.1118/1.2889720.