Effect of angle correction factors of MatriXX on VMAT dose verification
Yao Lei, Tan Caiyun, Liu Hancheng, Wang Haojie, Xia Bing
Department of Raiation Oncology,First Affiliated Hospital of China Medical University,Shenyang 110001,China (Yao L,Xia B);Beijing Allcure Medical Technology Co. Ltd. All Rights Reserved,Beijing 100010,China (Tan CY,Liu HCH,Wang HJ)
Download: PDF (1092 KB)   HTML (1 KB) 
Export: BibTeX | EndNote (RIS)      Supporting Info
Abstract  Objective To investigate the effect of the angle correction factors of ionization chamber array Matrixx on the dosimetric verification of volumetric modulated arc therapy plan. Methods The Matrixx (IBA) was put in the middle of the MultiCube Phantom. Computed tomography scan was performed and the results were sent to the Raystation treatment planning system (TPS). All data calculated from TPS and actually measured using MatriXX were gained under the following conditions:6 MV, 100 MU, 28 cm×28 cm radiation field. And the directions of beam were from 0°to 180°with an interval of 5°(except an interval of 1° from 85° to 95°). The angle correction factors (CF)(θ) cor of MatriXX were calculated by the calculated dose values from TPS and the measured values using MatriXX and then mirrored to the opposite side. CF (θ) cor values were compared with CF (θ) def values that were given by manufacturer. The comparison was made by paired t-test. Results CF (θ) cor and CF (θ) def were different, and the difference was significant between 85° and 95°, with a maximum of 7.5%(P<0.05). The CF (θ) cor correction method had a higher gamma pass rate than the CF (θ) def correction method, with a maximum of 17%(P<0.05). Conclusions The ionization chamber array MatriXX has an angular dependence and the angle correction should be performed. The angle correction factors have the individual characteristics and are different in every unit ionization chamber of the MatriXX. Therefore, it is necessary to do angle correction for every unit ionization chamber of the MatriXX and the possible errors caused by various factors should be considered in order to improve the accuracy rate and gamma pass rate of the plan verification.
Service
E-mail this article
Add to my bookshelf
Add to citation manager
E-mail Alert
RSS
Articles by authors
Yao Lei
Tan Caiyun
Liu Hancheng
Wang Haojie
Xia Bing
Key wordsIonization chamber matrix      Angle correction factor      Volumetric modulated arc radiotherapy      Plan verification     
Received: 07 March 2017     
Corresponding Authors: Xia Bing,Email:xia6581@sohu.com   
Cite this article:   
Yao Lei,Tan Caiyun,Liu Hancheng et al. Effect of angle correction factors of MatriXX on VMAT dose verification[J]. Chinese Journal of Radiation Oncology, 2017, 26(12): 1434-1437.