Application value of low-dose 4DCT scanning in simulation and target delineation for solitary pulmonary tumors
Shang Dongping, Luan Lijuan, Wang Lizhen, Sun Qingning, Yu Jinming, Yin Yong
Department of Radiation Oncology,Shandong Cancer Hospital and Institute, Shandong First Medical University and Shandong Academy of Medical Sciences,Ji′nan 250117,China
AbstractObjective To explore the feasibility of low-dose 4DCT scanning in simulation and target delineation for solitary pulmonary tumors (SPTs). Methods 23 patients with SPTs received 4DCT scanning simulation with the conventional scanning (CON), low voltage (LV), low current (LA) and low voltage combined with low current (LVA) in sequence. Based on the 4DCT images derived from the four sets of scanning parameters, the internal gross tumor volume (IGTVCON, IGTVV, IGTVA, IGTVVA) of SPTs were delineated and matched. Taking IGTVCON as reference, the tumor displacement and the centroid position of IGTVV, IGTVA and IGTVVA were compared with IGTVCON. The radiation doses under different scanning parameters were compared. Results The volumes of IGTVCON, IGTVV and IGTVA were (12.26±12.30) cm3,(12.21±12.16) cm3 and (11.87±11.70) cm3, respectively (P=0.337). IGTVVA was (11.34±11.07) cm3, significantly smaller than IGTVCON (P=0.005). There was no significant difference in the centroid positions of IGTVCON, IGTVV, IGTVA and IGTVVA in three directions (P=0.491, 0.360, 0.136). The Dice′s similarity coefficient (DSCVA) was significantly lower than DSCV and DSCA (P=0.004, 0.030). The tumor displacement measured by the four sets of 4DCT images was similar in the LR direction (P=0.470), and also in the AP direction (P=0.108). For the displacement in the SI direction, LVA scanning was smaller than CON (P=0.015). The radiation doses under four different scanning conditions were (397.0±140.3) mGy·cm,(175.0±61.8) mGy·cm,(264.8±95.3) mGy·cm and (116.8±41.2) mGy·cm, respectively (P<0.001). Conclusions LV or LA scanning exert no significant effect on the volume, centroid position of IGTV and the tumor displacement in 4DCT simulation for SPTs. The radiation dose that patients receive under LV and LV scanning is lower than that of CON. Consequently, LV or LA scanning is feasible in 4DCT simulation and target delineation for SPTs.
Shang Dongping,Luan Lijuan,Wang Lizhen et al. Application value of low-dose 4DCT scanning in simulation and target delineation for solitary pulmonary tumors[J]. Chinese Journal of Radiation Oncology, 2022, 31(2): 165-169.
Shang Dongping,Luan Lijuan,Wang Lizhen et al. Application value of low-dose 4DCT scanning in simulation and target delineation for solitary pulmonary tumors[J]. Chinese Journal of Radiation Oncology, 2022, 31(2): 165-169.
[1] Sun Q, Huang Y, Wang J, et al. Applying CT texture analysis to determine the prognostic value of subsolid nodules detected during low-dose CT screening[J]. Clin Radiol, 2019, 74(1):59-66. DOI:10.1016/j.crad.2018.07.103. [2] Kassem K, Shapiro M, Gorenstein L, et al. Evaluation of high-risk pulmonary nodules and pathologic correlation in patients enrolled in a low-dose computed tomography (LDCT) program[J]. J Thorac Dis, 2019, 11(4):1165-1169. DOI:10.21037/jtd.2019.04.31. [3] 李志刚,史鸿云,盖晓惠,等. 低剂量螺旋CT扫描对肺癌模拟定位应用价值[J]. 中华放射肿瘤学杂志,2013, 22(3):237-238. DOI:10.3760/cma.j.issn. 1004-4221.2013.03.019. Li ZG, Shi HY, Gai XH, et al. Practical value of low-dose CT scanning in simulation location of lung cancer[J]. Chin J Radiat Oncol, 2013, 22(3):237-238. DOI:10.3760/cma.j.issn.1004-4221.2013.03.019. [4] Henriksen OM, Larsen VA, Muhic A, et al. Simultaneous evaluation of brain tumour metabolism, structure and blood volume using[18F]-fluoroethyltyrosine (FET) PET/MRI:feasibility, agreement and initial experience[J]. Eur J Nucl Med Mol Imag, 2016, 43(1):103-112. DOI:10.1007/s00259-015-3183-6. [5] Coolens C, Bracken J, Driscoll B, et al. Dynamic volume vs. respiratory correlated 4DCT for motion assessment in radiation therapy simulation[J]. Med Phys, 2012, 39(5):2669-2681. DOI:10.1118/1.4704498 [6] Dou TH, Thomas DH, O′Connell DP, et al. A method for assessing ground-truth accuracy of the 5DCT technique[J]. Int J Radiat Oncol Biol Phys, 2015, 93(4):925-933. DOI:10.1016/j.ijrobp.2015.07.2272. [7] Wlfelschneider J, Seregni M, Fassi A, et al. Examination of a deformable motion model for respiratory movements and 4D dose calculations using different driving surrogates[J]. Med Phys, 2017, 44(6):2066-2076. DOI:10.1002/mp.12243. [8] Hugo Gd, Weiss E, Sleeman WC, et al. A longitudinal four-dimensional computed tomography and cone beam computed tomography dataset for image-guided radiation therapy research in lung cancer[J]. Med Phys, 2017, 44(2):762-771. DOI:10.1002/mp.12059. [9] 宋少娟,王巍,刘传亚. 多层CT螺距与辐射剂量的关系探讨[J]. 中华放射医学与防护杂志,2006, 26(5):526-528. DOI:10.3760/cma.j.issn.0254-5098.2006.05.034. Song SJ, Wang W, Liu CY. Relationship of radiation dose and spiral pitch for multi-slice CT system[J]. Chin J Radiol Med Protect, 2006, 26(5):526-528. DOI:10.3760/cma.j.issn.0254-5098.2006.05.034. [10] 张丽,于红,刘士远. 低剂量CT早期肺癌筛查的应用进展[J]. 国际医学放射学杂志,2011, 34(6):535-538. DOI:10.3784/j.issn.1674-1879.2011.06. Z0606. Zhang L, Yu H, Liu SY. Advances of early lung cancer screening with low-dose spiral CT[J]. Int J Med Radiol, 2011, 34(6):535-538. DOI:10.3784/j.issn.1674-1879.2011.06. Z0606. [11] 杜煜,时高峰,王琦,等. 低剂量CT对肺小结节大小定量测量影响的体模研究[J]. 中国临床医学影像杂志,2021, 32(1):47-51. DOI:10.12117/jccmi.2021. 01.012. Du Y, Shi GF, Wang Q, et al. Repeatability of small lung nodule measurement in low-dose CT:a phantom study[J]. J Chin Clin Med Imag, 2021, 32(1):47-51. DOI:10.12117/jccmi.2021.01.012. [12] Sarudis S, Karlsson-Hauer A, Nyman J, et al. Systematic evaluation of lung tumor motion using four-dimensional computed tomography[J]. Acta Oncol, 2017, 56(4):525-530. DOI:10.1080/0284186X.2016.1274049. [13] 高志鹏,王一更,李海洁,等. 实性肺结节低剂量CT影像组学图像质量控制的体模研究[J]. 中华放射学杂志,2020, 54(1):57-61. DOI:10.3760/cma.j.issn.1005-1201.2020.01.012. Gao ZP, Wang YG, Li HJ,et al. Phantom study in the quality control of low dose solid pulmonary nodule CT based on radiomics analysis[J]. Chin J Radiol, 2020, 54(1):57-61. DOI:10.3760/cma.j.issn.1005-1201. 2020.01.012. [14] Remy-Jardin M, Sobaszek A, Duhamel A, et al. Asbestos-related pleuropulmonary diseases:evaluation with low-dose four-detector row spiral CT[J]. Radiology, 2004, 233(1):182-190. DOI:10.1148/radiol.2331031133.