AbstractObjective To obtain the inter-fractional set-up errors of intensity-modulated radiotherapy (IMRT) of cervical cancer by cone-beam CT (CBCT), and to analyze the variations of the set-up errors on the cumulative dose deviation of the target volume. Methods A total of 48 patients with cervical cancer who underwent IMRT were enrolled in this study, and the set-up errors of 696 CBCTs were obtained. The set-up errors were input into the treatment planning system, and the cumulative set-up error dose was obtained by superposing the set-up errors dose. The deviation percentage was calculated by the deviation formula and the standard planning dose. Results The set-up errors caused the offset of isocenter distance by 0.58(0.36,0.80) cm. Different statistical differences were noted between the cumulative set-up error dose and the standard planning dose by WilCoxon test. All the dose deviations in the target volume were reduced, and the differential dose volume histogram (DVH) appeared negatively skewed, and the peak value was declined. The DVH diagram shifted to the left with an inverse S-curve and the slope was increased. The HIdeviation of the target volume from small to large were:CTV1,CTV2,GTV/CTV,CTV3,CTVn,CTVall,and GTVnd;The HIdeviation of the target volume were increased. Conclusions The effect of set-up errors in IMRT of cervical cancer upon the cumulative doses of the target volume significantly differs. The cumulative dose of the target volume is reduced, and the uniformity of the target volume dose becomes worse. The uncertainty of the inter-fractional position leads to an increase or decrease in the the fractional doses of the target volume. The biological effect on tumor cells and the tumor recurrence remains to be investigated. In IMRT of cervical cancer, the CBCT position calibration is required before each treatment to ensure the dose accuracy of each structure in the target volume.
Fund:Scientific Research Program of Health Industry in Gansu Province (GSWSKY-2015-29)
Cite this article:
Ouyang Shuigen,Tao Na,Liu Tingting et al. Effect of inter-fractional positional uncertainty on cumulative dose of target volume in intensity-modulated radiotherapy of cervical cancer[J]. Chinese Journal of Radiation Oncology, 2021, 30(8): 822-827.
Ouyang Shuigen,Tao Na,Liu Tingting et al. Effect of inter-fractional positional uncertainty on cumulative dose of target volume in intensity-modulated radiotherapy of cervical cancer[J]. Chinese Journal of Radiation Oncology, 2021, 30(8): 822-827.
[1] Mell LK, Roeske JC, Mundt AJ. A survey of intensity modulated radiation therapy use in the United States[J]. Cancer, 2003, 98(1):204-211. DOI:10.1002/cncr.11489. [2] Wagner A, Jhingran A, Gaffney D. Intensity modulated radiotherapy in gynecologic cancers:hope, hype or hyperbole?[J]. Gynecol Oncol, 2013, 130(1):229-236. DOI:10.1016/j.ygyno.2013.04.052. [3] 李玉成,程品晶,蒋璐,等. 基于EPID三维剂量重建在肿瘤患者中应用[J]. 中华放射肿瘤学杂志,2018,27(11):984-988. DOI:10.3760/cma.j.issn.1004-4221.2018.11.007. Li YC, Cheng PJ, Jiang L, et al. Application of EPID-based 3D dose reconstruction in cancer patients[J]. Chin J Radiat Oncol, 2018, 27(11):984-988. DOI:10.3760/cma.j.issn.1004-4221.2018.11.007. [4] 刘青,张宗恺,王雅棣,等. 图像引导对宫颈癌放疗中直肠、膀胱照射剂量的影响[J]. 中华放射肿瘤学杂志,2018, 27(1):68-73. DOI:10.3760/cma.j.issn.1004-4221.2018.01.014. Liu Q, Zhang ZK, Wang YD, et al. Effect of image guidance on dose to the rectum and bladder in radiotherapy for cervical cancer[J]. Chin J Radiat Oncol, 2018, 27(1):68-73. DOI:10.3760/cma.j.issn.1004-4221.2018.01.014. [5] International commission on radiation units and measurements. ICRU Report 50-prescribing,recording and reporting photon beam therapy[J]. Med Phys, 1994,21:833-844. DOI:10.1118/1.597396. [6] Michalski JM, Gay H, Jackson A, et al. Radiation dose-volume effects in radiation-induced rectal injury[J]. Int J Radiat Oncol Biol Phys, 2010, 76(3 suppl):S 123-129. DOI:10.1016/j.ijrobp.2009.03.078. [7] Hodapp N. The ICRU Report 83:prescribing.recording and reporting photon-beam intensity-modulated radiation therapy (IMRT)[J]. Strahlenther Onkol, 2012, 188(1):34-36. DOI:10.1093/jicru/ndq002. [8] 张健,岳海振,张艺宝,等. CT与CBCT图像配准范围选择的对比研究[J]. 中国医学物理学杂志,2015, 32(1):21-24. DOI:10.3969/j.issn.1005-202X.2015.01.006. Zhang J, Yue HZ, Zhang YB, et al. Registration accuracy comparison of CT and cone beam CT image based on different volumes of interested selection[J]. Chin J Med Phys, 2015, 32(1):21-24. DOI:10.3969/j.issn.1005-202X.2015.01.006. [9] Ahmad S, Vlachaki MT, Teslow TN, et al. Impact of setup uncertainty in the dosimetry of prostate and surrounding tissues in prostate cancer patients treated with Peacock/IMRT[J]. Med Dosim, 2005, 30(1):1-7. DOI:10.1016/j.meddos.2004.10.001. [10] 李明,修霞,王婵,等. 图像引导调强放疗摆位误差对老年前列腺癌患者受照剂量的影响[J]. 中华老年医学杂志,2015, 34(10):1107-1110. DOI:10.3760/cma.j.issn.0254-9026.2015.10.015. Li M, Xiu X, Wang C, et al. The impact of set-up errors on dose distribution during the image-guided intensity-modulated radiotherapy in elderly patients with prostate cancer[J]. Chin J Geriatrics, 2015, 34(10):1107-1110. DOI:10.3760/cma.j.issn.0254-9026.2015.10.015.